Press
ENTER
to accept this value. The digital display will revert to
Set Sensor
Span.
Press the joystick to the left to revert to the
Sensor Adjustment
menu.
6.2.1 Sensor Verification to Nitrogen
PureAire recommends challenging the O
2
monitor with nitrogen every 6 to 12
months. The monitors dust filter has a ¼” male tube fitting designed for
connecting sample tubing from a Nitrogen cylinder. Expose the O
2
cell to N
2
using the On-Demand regulator. The reading will drop off to 1% or below in
less than one minute when the O
2
sensor is exposed to pure N
2
. The system will
recover to 20.9% when the nitrogen is removed.
NOTE: Please Remove the water/dust filter from the monitor
before connecting the gas cylinder. The span gas cylinder must
be connected directly to the monitor as shown in the following
photo on pg. 45.