PureAire Monitoring Systems, Inc.
33 |
P a g e
NOTE:
The O
2
system must have the audio alarm option module installed to access
this menu. If this option is installed the display will indicate N/A, (not available)
NOTE: The optional built-in horn is designed to operate in only one alarm
mode. It will activate in either a decreasing alarm or an increasing
alarm mode only. The horn activation is immediate any time an
alarm threshold is exceeded.
To change this value, push the joystick right to display the relay state. The display will
Indicate
INVERT.
Pushing the joystick down will change the relay state from INVERT
to NORMAL. Press
ENTER
to accept the value.
Auto Mode -
The auto mode is used when you wish the horn to activate at the same
time the relays activate. In the Normal or Inverted Mode, the horn
immediately activates any time the alarm thresholds are exceeded. To
activate the horn when the relays activate, choose the AUTO mode.
After entering the relay state the display will default back to the
Set Audio Alarm
Polarity
menu. The display will scroll the following:
5.5.4 Set Latching
This is the menu at which to adjust the relay alarm state for the two gas alarm relays and
the individual instrument fault relay. The selection permits setting the relays to a
latching or non-latching state. In a latching state, the relay will remain activated until
the user manually selects the Enter Key. In a non-latching state, the alarm relay will
automatically reset once the gas concentration has returned to 20.9% for oxygen.
INVERT
20.9%
NORMAL
20.9%
..Set Audio Alarm Polarity...
20.9%
..Set Audio Alarm Polarity...
20.9%
AUTO
20.9%