5.3 Sensor Calibration
NOTE:
This procedure should be performed in a gas free atmosphere at
normal room temperature (68°F; 20°C).
The
Air Check
Lite
requires periodic calibration with the appropriate standard
gas. Calibration should be performed whenever:
The membrane or electrolyte is replaced;
The entire sensor is replaced;
Six months has passed without membrane, electrolyte, or sensor
replacement.
The PureAire Calibration Kit (optional) is recommended for calibration.
5.3.1 Gas Generation and Calibrating Kits
For generating a calibrating gas, the following PureAire Gas Generation Kits are recommended.
They are available by separate order.
Calibration Kit Model
K-I
PH
3
K-II
Cl
2
K-III
HCN
K-IV
SO
3
K-V
H
2
S
K-VI
NH
3
K-VII
HCl
K-VIII
HF
Gas generation tubes
1 box
1 box
1 box
1 box
Gas generation solution (10 ml)
2
bottles
2
bottles
2
bottles
2
bottles
Soft paper towels
1 pack
1 pack
1 pack
1 pack
1 pack
Gas detection tube
1 box
1 box
1 box
1 box
1 box
1 box
1 box
1 box
Gas sampling bags (2 liter)
2
2
2
2
2
2
2
2
Gas sampling pump
No. 800
1
1
1
1
1
1
1
1
Double bellows
1
1
1
1
1
1
1
1
Teflon
tube (
6
4)
1 m
1 m
1 m
1 m
1 m
1 m
1 m
1 m
Calibration Cap (optional)
1
1
1
1
1
1
1
Mini-pump MP-05 (optional)
1
1
1
1
1
1
1
Instruction manual
1
1
1
1
1
1
1
1
Case
1
1
1
1
1
1
1
1
Silicon tube (
8
4)
1 m
1 m
1 m
1 m
1 m
1 m
1 m
1 m
Reagent 1
1 bottle
1 bottle 1 bottle 1 bottle
Reagent 2
1 bottle
1 bottle 1 bottle 1 bottle
Dispenser bottle (100 ml)
1 bottle
1 bottle 1 bottle 1 bottle
Summary of Contents for Air Check Lite
Page 26: ...PureAire Monitoring Systems Inc 5 25 07 23 6 Drawings ...
Page 28: ...PureAire Monitoring Systems Inc 5 25 07 25 DQS cell Mounted to AirCheck Lite ...
Page 29: ...PureAire Monitoring Systems Inc 5 25 07 26 Membrane ...
Page 30: ...PureAire Monitoring Systems Inc 5 25 07 27 24VDC 3 wire is remote up to 1000 meters ...