PureAire Monitoring Systems, Inc.
5/25/07
18
11.
Wipe the pressure compensation port and its surrounding area with a dry
paper tissue.
Paper Tissue
Pressure Compensation
Port
12.
Insert the pressure compensation plug into the port and cover with a new
milli seal.
Pressure
Compensation
Screw
Milli Seal
13.
Reinstall the sensor as outlined in section 5.1.2.
NOTE:
When storing the sensor cell never store the sensor cell horizontal with the Pressure
Compensation Screw positioned down or store the sensor cell upside down. This can cause the
electrolyte to leak from the sensor cell pressure compensation screw.
Pressure Compensation Port
Pressure Compensation Screw