PureAire Monitoring Systems, Inc.
32
Pushing the joystick up increases the value and pushing the joystick down decreases the
value. Press
ENTER
to accept the value.
After entering the relay state, the display will default back to the
Set Relay 2 Alarm
Threshold
Menu. The display will scroll the following:
Push the joystick down to access the next sub menu;
Set Audio Alarm Threshold,
will scroll on the digital display.
This is the gas concentration at which the instrument’s audio alarm will be activated.
To change the displayed value, push the joystick to the right to display the second level
alarm setting. The display will indicate a value between 00.0% and 25.5%.
Pushing the joystick up increases the value and pushing the joystick down decreases the
value. Press
ENTER
to accept the value.
NOTE:
The O
2
system must have the audio alarm option module installed to access
this menu. If this option is installed the display will indicate N/A, (not available)
NOTE:
The audio can be set into only one alarm level. You can choose between
alarm level 1 or alarm level 2 or set a completely different setting.
After entering the relay state, the display will default back to the
Set Audio Threshold.
Menu. The display will scroll the following:
18.0%
20.9%
..Set Audio Alarm Threshold…
20.9%
19.5%
20.9%
..Set Audio Alarm Threshold…
20.9%
..Set Relay 2 Alarm Threshold…
20.9%
NOTE:
The
menus
highlighted in
light gray are
only used for
O2 monitors
with the Alarm
Relay Board
They are
displayed on
the O
2
monitor
and their
function is
disabled.