
Because the trip fault detectors (P) do not operate for external
faults, underreaching transfer trip systems do not require external
fault-clearing coordination circuits (transient blocking) and are,
therefore, inherently simpler than any of the other schemes. You
obtain maximum security if you use additional permissive fault
detectors. These schemes also provide minimum operating times
for many faults that are tripped directly, without using the
channel.
December 1996
Page 2–5
Chapter 2. Applications and Ordering Information
2
Figure 2–4. Basic Operation of the Dual
Phase Comparison Pilot Relaying System.
2.1.4
Dual Phase
Comparison
Unblocking
Systems
Dual comparison systems require a
duplex channel: one frequency for
each line terminal. The TCF–10B
frequency-shift channel equipment
is available for this purpose;
normally used in an unblocking
system. Continuous channel moni-
toring is also provided, because
either a mark or space carrier
signal is always transmitted.
The transmitter is keyed to its
mark frequency when the square
wave from the filter goes positive,
and is keyed to its space frequency
when the square wave is at zero.
There are two outputs at the
receiver: the “mark” output is a
square wave that goes positive
when a mark frequency is
received; the “space” output goes
positive when a space frequency is
received.
The basic operation of the Dual
Phase Comparison system is
shown in Figure 2-4. For internal
faults, the single phase outputs of
the sequence current networks are
essentially in phase, although such
output represents currents 180°
apart in the power system. The
network output goes through a
squaring amplifier that keys the
frequency shift transmitter. An
adjustable delay circuit delays the
local square wave by a time equal
to the channel delay time.
The network output is then used to
develop two complementary
square waves. One wave, which
has a positive state during the
positive half-cycle of the sequence
current network, is compared with
the receiver’s mark output. The
Summary of Contents for TCF-10B
Page 2: ......
Page 32: ...Page 1 18 April 1997 TCF 10B System Manual Technologies Inc USER NOTES ...
Page 49: ...December 1996 Foigure 2 15 TCF 10B Catalog Numbers Module Style Numbers 1355D19 Page 2 17 ...
Page 57: ...Page 3 8 April 1997 TCF 10B System Manual Technologies Inc USER NOTES ...
Page 58: ...April 1997 Figure 3 3 TC 10B TCF 10B Mechanical Outline Drawing 1354D48 Page 3 9 ...
Page 62: ...Page 4 2 January 1996 TCF 10B System Manual Technologies Inc Figure 4 1 Extender Board ...
Page 88: ...Page 6 12 April 1997 TCF 10B System Manual Technologies Inc 8 RECEIVER LOGIC ...
Page 89: ...April 1997 Page 6 13 Chapter 6 Routine Adjustment Procedures 6 ...
Page 95: ...April 1997 Figure 7 1 TCF 10B Functional Block Diagram 1354D13 Page 7 5 ...
Page 117: ...Page 10 12 January 1996 TCF 10B System Manual Technologies Inc USER NOTES ...
Page 123: ...Page 11 6 Figure 11 2 TCF 10B Transmitter Schematic 1355D71 April 1997 ...
Page 129: ...Page 11 12 April 1997 TCF 10B System Manual Technologies Inc USER NOTES ...
Page 143: ...Page 13 6 April 1997 TCF 10B System Manual Technologies Inc USER NOTES ...
Page 167: ...Page 14 24 April 1996 TCF 10B System Manual Technologies Inc USER NOTES ...
Page 189: ...April 1997 Figure 16 7 2 Frequency Directional Comparison Functional Block Diagram Page 16 7 ...
Page 190: ...Page 16 8 Figure 16 8 3 Frequency Directional Comparison Functional Block Diagram April 1997 ...
Page 204: ...April 1997 Figure 16 11 TCF 10B Receiver Logic Schematic CF30RXLMN Sheet 1 of 3 Page 16 21 ...
Page 205: ...Page 16 22 Figure 16 12 TCF 10B Receiver Logic Schematic CF30RXLMN Sheet 2 of 3 April 1997 ...
Page 206: ...April 1997 Figure 16 13 TCF 10B Receiver Logic Schematic CF30RXLMN Sheet 3 of 3 Page 16 23 ...
Page 214: ...Page 17 8 January 1996 TCF 10B System Manual Technologies Inc USER NOTES ...
Page 232: ...Page 18 18 April 1997 TCF 10B System Manual Technologies Inc USER NOTES ...