DIAMENTOR E2
Electromagnetic Compatibility (EMC) according to IEC 60601-1-2 / Special Adjustments
30
D651.131.00/05 en
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The DIAMENTOR E2 is intended for use in the electromagnetic environment specified below.
The customer or the user of the DIAMENTOR E2 should assure that it is used in such an environment.
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment – Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
V
1
= 10 V
rms
2
E
1
= 3 V/m
Portable and mobile RF communications equipment
are used no closer to any part of the
DIAMENTOR E2, including cables, than the recom-
mended separation distance calculated from the
equation appropriate for the frequency of the trans-
mitter.
Recommended Separation Distance
d =
P
V
1
5
.
3
=
1.2
P
3
d =
P
E
1
5
.
3
=
1.2
P
80 MHz to 800 MHz
d =
P
E
1
7
=
2.3
P
800 MHz to 2.5 GHz
where P is the power rating of the transmitter in
watts (W) according to the transmitter manufacturer
and d is the recommended separation distance in
meters (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey,
a)
are less
than the compliance level in each frequency range.
b)
Interference may occur in the vicinity of equipment
marked with the following symbol:
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If the
measured field strength exceeds the RF compliance level above, observe the DIAMENTOR E2 to verify normal opera-
tion in each use location. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the DIAMENTOR E2.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and re-
flection from structures, objects and people.
2
different test level from product standard IEC 60580 (IEC 61000-4-6) was used
3
calculation with test level of IEC 60601-1-2