MDM300 User Manual
32
97122 Issue 16.2 June 2022
OPERATION
The display shows:
#
Indicates a log number for each set of log points
INT
Value of primary display parameter
EXT
Value of secondary display parameter
ST
Status column which reports the status of the instrument at each log
point in hexadecimal code.
The status codes are explained in Appendix B.
•
FILE
File name
•
INT
Primary display parameter being logged
•
EXT
Secondary display parameter being logged
•
INTVAL
Logging interval
•
D#
Current date
in YYMMDD format
•
T#
Current time
in HHMMSS format
NOTE: The chart and datalogging are independent functions; therefore the
parameter displayed on the chart may not necessarily reflect what is being
logged.
3.4.11 CALIBRATION
Select calibration
MDM300 Calibration Correction
Easidew Offset Adjustment
Easidew One Point Adjustment
Figure 28
CALIBRATION Page
Entry to the CALIBRATION Menu is protected by a passcode (
4876
). The passcode is
entered in a similar manner to that described in Section 3.3.1.
This page gives access to the internal sensor’s calibration look-up table and provides the
facility to select and change any single point, or number of points. It also allows an
offset to be applied to a connected Easidew transmitter.
Section 6 details the use of the calibration routines.
Summary of Contents for MICHELL Instruments MDM300
Page 1: ...MDM300 Advanced Dew Point Hygrometer User Manual 97122 Issue 16 2 June 2022 ...
Page 19: ...MDM300 User Manual Michell Instruments 11 INSTALLATION 1 4 3 6 5 2 Figure 6 Accessories ...
Page 61: ...MDM300 User Manual Michell Instruments 53 APPENDIX B Appendix B Datalog Status Display ...
Page 64: ...MDM300 User Manual 56 97122 Issue 16 2 June 2022 APPENDIX C Appendix C FCC Declaration ...