INSTALLATION
2.6.2 Set-Up External Temperature Transmitter Parameters
After adding a temperature transmitter, the following parameters must be defined.
Switch on the instrument and initially set the required
DP/TEMP
units to either
°C, °F
or K
from the Settings Menu (refer to Section 3.4.1).
Using the
EXTERNAL
option from the SET-UP Menu, set-up the instrument for an
external temperature transmitter input as follows (refer to Section 3.4.4).
EXTERNAL TYPE
Select
TEMP
EXTERNAL ZERO
Enter Zero point (temperature represented by 4 mA input)
EXTERNAL SPAN
Enter Span (maximum temperature represented by 20 mA input)
After completion of the above settings, return to the Main Display and the instrument
will now display temperature in the secondary display area and the selected units.
An example is shown below:
Dewpoint
External
°C
12.7
6.9
°C Temp
Figure 11
Typical External Temperature Display
The measured temperature will be used to calculate Relative Humidity.
2.6.3 Set-Up External Dp (Easidew) Transmitter Parameters
After adding an external dew-point transmitter, the following parameters must be
defined. Switch on the instrument and initially set the required
DP/TEMP
units to either
°C, °F
or
K
from the Settings Menu (refer to Section 3.4.1).
Using the
EXTERNAL
option from the SET-UP Menu, set-up the instrument for an
Easidew transmitter input as follows (refer to Section 3.4.4).
EXTERNAL TYPE
Select
EASIDEW
EXTERNAL ZERO
-100 °C (
DP TEMP
represented by 4 mA input - can be reset)
EXTERNAL SPAN
20 °C (
DP TEMP
represented by 20 mA input - can be reset)
After completion of the above settings, return to the Main Display and the instrument
will now display pressure in the secondary display area and the selected units.
An example is shown below:
Calc dp at pressure
External
°C
-25.7
-26.9
°C dp
Figure 12
Typical External Dew-Point Display
Summary of Contents for MICHELL Instruments MDM300
Page 1: ...MDM300 Advanced Dew Point Hygrometer User Manual 97122 Issue 16 2 June 2022 ...
Page 19: ...MDM300 User Manual Michell Instruments 11 INSTALLATION 1 4 3 6 5 2 Figure 6 Accessories ...
Page 61: ...MDM300 User Manual Michell Instruments 53 APPENDIX B Appendix B Datalog Status Display ...
Page 64: ...MDM300 User Manual 56 97122 Issue 16 2 June 2022 APPENDIX C Appendix C FCC Declaration ...