Do not insert metal objects such as a pin or a wire into any gap or any outlet on the device.
❚
Do not touch the patient and external electrical devices at the same time.
❚
Always the latest command will be executed, regardless of whether it came via interface or from the UserTerminal during one of
the four modes. Only stop command has higher priority and cannot be overwritten.
❚
WARNING: To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
❚
WARNING: Do not use portable high frequency communication devices in the patient environment (see “position of patient and
user”). Disregard can cause loss of performance.
❚
Free standing equipment has to be installed on a stable and levelled base.
❚
Choose proper floor, clothing and humidity, in order to prevent electrostatic discharge (also see technical data).
❚
Do not use the device without instruction by authorized personnel acc. to the instruction protocol.
❚
Regard safety area behind device of 2.0 m x width of treadmill.
❚
Do not use the device with children <12 months.
❚
Exclude access of unsupervised children (< 14 years) onto or near any parts of the device (incl. accessories, packaging, lubrication
and service material).
❚
In case of application with children (> 1, < 14 years) permanent observation of the patient by medical staff is obligatory.
❚
Animals must not be in the same room with the device.
❚
The automatic modes must only be performed on the prescription of the medical doctor.
❚
Operator and patient have to be aware of automatic load changes during profile, cardio and test mode.
❚
During stress tests a medical doctor has to be available at any time.
❚
Unmeant trapping hazards: Take off ties, scarfs or other clothes that may be trapped. Secure long hair and ribbons during
maintenance and training in order to prevent being captured in trapping zones.
❚
Perform a daily visual inspection (see chapter “maintenance”).
❚
Obey the maintenance intervals claimed in chapter “maintenance”.
❚
Obey the competences claimed in chapter “maintenance”.
❚
A second person has to be present during maintenance.
❚
In case of any visible or assumed defects or malfunctions (of the device, accessories, software, etc.), unplug device, exclude
reconnection, mark clearly and inform h/p/cosmos service personnel via telephone and writing.
❚
In case of any visible or assumed wear and tear (of the device, accessories, labels, etc.), unplug device, exclude reconnection,
mark clearly and inform h/p/cosmos service personnel via telephone and writing.
❚
In case of any fluid entering into the device, unplug device, exclude reconnection, mark clearly and inform h/p/cosmos service
personnel via telephone and writing.
❚
Do not modify the device, configurations, accessories or software in any way.
❚
Do not connect any devices, accessories or software, not listed in “accessories / compatible devices”.
❚
Disinfect the device before and after every treatment.
❚
Disconnect the device and all accessories from mains power supply before cleaning or disinfection.