10.2 Gas flow mode (optional)
A3480
Setpoint
Standard
Operation Mode
Constant
Gasflow
Constant
ozone concentration
Gasflow Min.
Gasflow Max.
Reset
Fig. 12: Gas flow mode (optional)
The following operating modes can be selected for the gas flow
control valve:
n
‘ On’ ‘Operation Mode’ ‘Constant Gasflow’
n
‘ On’ ‘Operation Mode’ ‘Constant ozone concentration’
10.2.1
‘On’ ‘Operation Mode’ ‘Flow’
INFORMATION! The value range for the possible inputs depends
on the type of system.
A3480
Setpoint
Standard
Operation Mode
Constant
Gasflow
Constant
ozone concentration
Gasflow Min.
Gasflow Max.
Reset
Fig. 13: ‘Operation Mode’ ‘Gas flow’
The control valve is in operation. The gas flow is regulated. You
can enter the
‘Constant gas flow’ setpoint in the ‘Setpoint’ field.
The
‘Constant gas flow’ field indicates the current gas flow in
Nm
3
/h (standard cubic metres per hour). Operation with a largely
constant gas flow is possible in this operating mode. In this way,
pressure changes in the raw water, which modify the gas flow in
normal operation, can be largely compensated for.
10.2.2
‘On’ ‘Operation Mode’ ‘Constant ozone concentration’
INFORMATION! The value range for the possible inputs depends
on the type of system.
Operating modes and settings
40