ProMinent gamma/ X Operating Instructions Manual Download Page 61

1.

The pump control stops the normal metering operation - the 

"Stop Pause" symbol appears on the LCD screen.

2.

Only with bleed relay: After 1 s, the pump control opens the bleed

valve on the liquid end (via the bleed relay and the solenoid valve).

3.

1 s later, the pump starts to run at its maximum possible stroke rate

(as during priming) - the   "Air lock" symbol appears on the LCD

screen instead of the "Stop Pause" symbol.

4.

The pump operates in this way for the total set time.

5.

Once the set time has elapsed, the pump stops – the "Stop Pause"

 symbol appears on the LCD screen again.

6.

Only with bleed relay: After 1 s the pump control closes the liquid

end bleed valve.

7.

After 1 s the 

 "Stop Pause" symbol disappears and the pump

recommences normal operation.

If the pump was in the "Stop" state at the time of triggering (  

[STOP/START]

, Pause, Error), the start of the bleed procedure is delayed

- until this state has been cleared.
If the pump switched to "Stop" state during the bleed procedure, the pump

control jumps immediately to phases 5 and 6 - see above. This definitely

cancels the bleed procedure. As soon as the "Stop" state is cleared, the

bleed procedure starts from the beginning.

11.2.9 

‘Priming Time’

‘Menu / Information

 

è

 Settings

 

è

 Priming time

 

è

 ...’

In the 

‘Priming time’

 menu, you can select how long the metering pump is

to prime once    

[Priming]

 has been pressed.

In operation, the stroke length can be changed more easily

by using the "Priming time" display:

Press    [Priming] - the pump starts to prime.

Turn the [Clickwheel] to set the priming time.

Press the [Clickwheel] to save the priming time.

11.2.10 

‘Time’

‘Menu / Information

 

è

 Settings

 

è

 Time

 

è

 ...’

In the 

‘Time’

 menu you can set the time.

1.

Use the dial to adjust a figure.

2.

Use    

[Priming]

 to move to the next figure.

Under 

‘Auto. Summer time’

 you can select whether you wish changeover

to 

‘Summer time’

.

You can also input when the pump is to change over and back to and from

‘Summer time’

.

Sequence of the bleed procedure (auto‐

matic):

Set Up / 

‘Menu’

 

 

 

61

Summary of Contents for gamma/ X

Page 1: ...ons 2006 42 EC Part no 984586 BA G 044 04 15 EN Please carefully read these operating instructions before use Do not discard The operator shall be liable for any damage caused by installation or opera...

Page 2: ...level 1 Menu level 2 Menu level Menu paths Software interface text Information This provides important information relating to the correct operation of the unit or is intended to make your work easier...

Page 3: ...1 2 Installation of Metering Pumps With Bleed Valve 31 8 1 3 Installation of Metering Pumps With Self bleeding SEK Type 31 8 2 Basic installation notes 32 9 Installation Electrical 34 9 1 Supply volt...

Page 4: ...ical Data 78 17 1 Performance Data 78 17 2 Accuracy 79 17 2 1 Standard Liquid End 79 17 2 2 Self Bleeding Liquid End 79 17 3 Viscosity 79 17 4 Material specifications 80 17 5 Electrical Data 80 17 6 T...

Page 5: ...valve spring for more high viscosity media 9 Self bleeding with bypass SEK Hydraulic connector 0 Standard connection in line with technical data 5 Connector for 12 6 hose suction side standard 9 Conne...

Page 6: ...n valve 2 m suction line 5 m metering line 4 Multifunctional valve and accessories Control version 0 Manual external contact with pulse control 3 Manual external contact with pulse control analogue 0...

Page 7: ...n for simple adjustment with volume proportional metering tasks n Stroke rate adjustment in 1 stroke hour increments from 0 12 000 strokes h n Electronic stroke length adjustment continuous from 0 100...

Page 8: ...ing instructions n Observe the general limitations with regard to viscosity limits chem ical resistance and density see also ProMinent resistance list in the Product Catalogue or at www prominent com...

Page 9: ...when working on the pump Take appropriate protective measures before working on the pump e g safety glasses safety gloves Observe the safety data sheet for the feed chemical Drain and flush the liqui...

Page 10: ...le feed chemical can damage the wetted parts of the pump Take into account the resistance of the wetted materials when selecting the feed chemical see the ProMinent Product Catalogue or visit our home...

Page 11: ...sport unpacking Instructed person Assembly Technical personnel service Planning the hydraulic installation Qualified personnel who have a thorough knowledge of metering pumps Hydraulic installation Te...

Page 12: ...instructed and if required trained in the tasks assigned to him her and possible dan gers that could result from improper behaviour as well as having been instructed in the required protective equipm...

Page 13: ...page CAUTION Danger of material damage The device can be damaged by incorrect or improper storage or transportation The unit should only be stored or transported in a well packaged state preferably in...

Page 14: ...2 3 4 5 6 7 1 2 3 4 5 6 7 8 2 3 1 6 7 A B C P_G_0053_SW Fig 3 A Liquid end with PV bleed valve B Liquid end with NP bleed valve C Self bleeding liquid end SEK 1 Discharge valve 2 Backplate 3 Dosing h...

Page 15: ...tor yellow 9 Operating indicator green 10 Diaphragm rupture indicator terminal 11 External control terminal 12 Dosing monitor terminal 13 Level switch terminal 14 Slot for relay and optional modules 5...

Page 16: ...mA i 4 mA i 4 mA B0605 a b i 4 mA Fig 6 a Continuous display with warning message b Continuous display with fault message Explanation of the symbols in the following tables The above Figure Part a sho...

Page 17: ...and BATCH operating modes The auxiliary function Stroke memory has been set The pump is in ANALOGUE operating mode The Curve linear type of processing is set The pump is in ANALOGUE operating mode Th...

Page 18: ...ibrated state in l or l h or in gal or gal h 5 2 2 Key Functions Key Application In the continuous displays In the menu Back press Move back to the previous menu point or a continuous display without...

Page 19: ...Refer to the Set up Basics chapter to adjust figures Overview of Equipment and Control Elements 19...

Page 20: ...the flow to be precisely matched to the requirements of the particular applica tion This ensures that the user can set the optimum discharge stroke for his application as required Pos Discharge stroke...

Page 21: ...naccurate metering with high viscosity feed chemicals namely the incomplete filling of the liquid end With gaseous feed chemicals the slow suction stroke prevents cavitation and conse quently increase...

Page 22: ...epressurised state due to the integral back pressure valve 6 5 Operating Modes Operating modes are selected via the Operating modes menu Refer to the Hierarchy of Operating Modes Functions and Fault S...

Page 23: ...number of defective strokes after which the pump is switched off can be set in the Settings menu Information about the liquid level in the dosing tank is reported to the pump A two stage level switch...

Page 24: ...icator red A fault message is pending undefined operating status Warning indicator yellow A warning message is pending Operating indicator green The pump is ready for operation At each stroke Stroke r...

Page 25: ...of the liquid end are not vertical they cannot close correctly Ensure that the suction and discharge valves are upright with self bleeding liquid ends the bleed valve Several metering pumps can be ar...

Page 26: ...olts and washers for fixing the assembly foot Bolt Type Size Washer DELTA PT bolt 50 WN5412 5452 A5 3 DIN 125 PT bolt 50 WN1441 1411 KA B Flat head bolt M5 DIN EN ISO 7045 A5 3 DIN 125 Allen bolt M5 D...

Page 27: ...ing of feed chemical spraying around Feed chemical can spray out of the hydraulic components if they are manipulated or opened due to pressure in the liquid end and adjacent parts of the system Discon...

Page 28: ...f all hydraulic components please refer to the product specific operating instructions and system documentation Never allow the metering pump to run against a closed shut off device Install a relief v...

Page 29: ...zle 4 and widen if necessary Ensure that the O ring and flat seal 5 is properly fitted to the valve 6 Never re use used PTFE seals An installation sealed in this way is not watertight This type of sea...

Page 30: ...Front clamp ring 5 Valve CAUTION Warning of feed chemical spraying around Connections can come loose in the event that hose lines are installed incorrectly on stainless steel valves Only use PE or PT...

Page 31: ...riming pressure on the suction end is at least equal to the return line pressure Back pressure in the return line restricts the bleeding function In contrast it is possible to operate the pump with ba...

Page 32: ...g hydraulic components Hydraulic components can rupture if the maximum permis sible operating pressure is exceeded Never allow the metering pump to run against a closed shut off device With metering p...

Page 33: ...stallation 1 Main line 2 Storage tank Symbol Explanation Symbol Explanation Metering pump Foot valve with filter meshes Injection valve Level switch Multifunctional valve Manometer Legend for hydrauli...

Page 34: ...etely installed electrical options can allow moisture into the inside of the housing Fit appropriate modules into the slot on the front of the pump or use the original blank cover to seal it in a leak...

Page 35: ...control Terminal The external control terminal is a 5 pole panel terminal It is compatible with 2 and 4 pole cables Only use a 5 pole cable with the Auxiliary capacity Auxiliary frequency and mA inpu...

Page 36: ...same time pin 1 and pin 4 must also be connected to each other The pump capacity and or stroke rate can be controlled by a current signal The current signal is connected between pin 3 and pin 4 In add...

Page 37: ...g monitor Electrical interface Data Value Unit Voltage with open contacts 5 V Input resistance 10 k Control via n potential free contact load 0 5 mA at 5 V or Pin Function 4 wire cable 1 Power supply...

Page 38: ...act 230 V 8 A X 4 fault indicating relay Pacing relay N O N O 24 V 24 V 100 mA 100 mA X X C fault indicating relay 4 20 mA output N O 24 V 100 mA X Relay type switches in the event of Relay type Level...

Page 39: ...green NC normally closed red 4 brown C common black 9 2 5 3 Output for Other Relays Identity code 4 A fault indicating and a pacing relay can optionally be ordered refer to ordering information in th...

Page 40: ...us relay identity code C A relay combined with a current output can optionally be ordered The relay either switches off the pump as a fault indicating relay in the event of a fault on the pump and wit...

Page 41: ...idual voltage at Ioff max 1 A 0 4 V Maximum current 100 mA Maximum voltage 24 VDC Pacing pulse duration approx 100 ms To pin VDE cable Contact Relay 1 yellow Current output 4 green Current output 3 wh...

Page 42: ...e Makro operating mode Information Menu German English Language Language German Menu Information Path derived Language Language English German Save Menu Information Path operating instructions Languag...

Page 43: ...Changing from one figure to its initial figures b Changing the figure c Returning from the last figure to the complete figure to correct a wrong figure for example0 Changing a complete number Turn the...

Page 44: ...ontinuous display shows different information which cannot be adjusted in the secondary display see overview Continuous displays and secondary displays in the appendix You can access secondary display...

Page 45: ...s follows 1 Information 2 Settings 3 Service 4 Language 11 1 Information Menu Information The Information provides information on your pump and certain parame ters and counters The number and type can...

Page 46: ...he maximum stroke rate can be set in Contact operating mode It should normally be set to 12 000 strokes hour You can also activate the Memory function extension memory identi fier When Memory is activ...

Page 47: ...1 stroke is executed after 2 pulses With a factor of 0 1 1 stroke is executed after 10 pulses With a factor of 0 75 1 stroke is executed once after 2 pulses then 1 stroke is executed twice after 1 pul...

Page 48: ...not be processed up to the maximum capacity of the stroke memory of 4 200 000 000 strokes If this maximum capacity is exceeded the pump goes into fault mode CAUTION The pump maintains its stroke rate...

Page 49: ...B0088 Fig 31 Frequency current diagram for Linear curve Plot a diagram similar to the one above with values for I1 F1 and I2 F2 so that you can set the pump as desired The smallest processable differe...

Page 50: ...g volume or dosing time instead of stroke length and stroke rate The other selection points will also appear in the operating menu refer to the chapter Operating Menu for gamma X complete in the appen...

Page 51: ...r application This ensures that the user can set the optimum discharge stroke for his application as required Pos Discharge stroke Application A optimum For maximum accuracy when metering and the very...

Page 52: ...roke It is possible to selectively also slow the suction stroke with all these metering profiles for the discharge stroke see In this way it is possible to prevent the main cause of inaccurate meterin...

Page 53: ...m stroke rate Strokes min Normal 200 HV1 160 HV2 120 HV3 80 11 2 4 3 Pressure Stage Menu Information Settings Metering Pressure stage You can use the programmable function Pressure stage to reduce the...

Page 54: ...Monitoring 11 2 4 4 1 Air lock Menu Information Settings Metering Monitoring Air lock Should a message and the Air lock symbol appear this may mean that there is gas in the liquid end if Warning or E...

Page 55: ...hemical t1 t2 Q1 Q2 P_G_0071_SW 1 Use the Clickwheel to scroll through the continuous display to check whether litres or gallons have been selected 2 If the incorrect volume unit has been selected cor...

Page 56: ...e pump returns to the Menu Information Settings menu 11 2 6 System Menu Information Settings System The System menu splits into the following sub menus 1 Dosing head 2 Volume unit 3 Pressure unit 11 2...

Page 57: ...Information Settings Inputs Outputs Auxiliary capacity Auxiliary frequency The programmable function Auxiliary capacity Auxiliary frequency permits the switch over to an additional capacity stroke rat...

Page 58: ...e Pause Metering Batch The relay changes its state as soon as a batch has been processed Stroke rate The relay switches at every stroke Option The relay has the option of the module plugged in e g the...

Page 59: ...e 11 2 7 6 Flow Control Menu Information Settings Inputs Outputs Flow control The setting options for the Flow control function only exists if a flow control has been electrically installed The symbol...

Page 60: ...y Cycle and Duration 3 By the Air lock internal signal 4 If one of the two signals occurs Both Detailed explanation 1 If Off has been selected in the menu this function is deactivated 2 If Periodicall...

Page 61: ...ed until this state has been cleared If the pump switched to Stop state during the bleed procedure the pump control jumps immediately to phases 5 and 6 see above This definitely cancels the bleed proc...

Page 62: ...t Point none Lock menu X Lock all X X STOP STOP START START 1 Setting mode Continuous display Stop start pump Priming Start batch only in Batch operating mode Acknowledging errors Check adjustable val...

Page 63: ...the last cali bration or the last deletion 11 3 4 Diaphragm Replacement Menu Information Service Diaphragm replacement You can move the slide rod into the replacement position here with To replacement...

Page 64: ...s STOP START again Press Priming Turn the Clickwheel to extend of shorten the priming time during priming In Batch operating mode Press the Clickwheel in the Push contin uous display Press the Clickwh...

Page 65: ...TART START 1 Setting mode Continuous display Stop start pump Priming Start batch only in Batch operating mode Acknowledging errors Check adjustable values 2 Changing directly changeable variables B059...

Page 66: ...ead to problems when pumping Use only original spare parts Use the correct spare parts kits In the event of doubt refer to the exploded views and ordering information in the appendix Interval Maintena...

Page 67: ...e bleed valve is tight n Check the discharge and bypass line for kinks n Check that the bleed valve is operating correctly Under normal loading approx 30 of continuous operation Under heavy loading e...

Page 68: ...ed during repair work Protect yourself against the feed chemical in case it is hazardous Read the safety data sheet on the feed chemical CAUTION Warning of feed chemical spraying around Feed chemical...

Page 69: ...14 1 Cleaning Valves Warning of faulty operation Refer to the exploded drawings in the appendix when working on the unit Repairs 69...

Page 70: ...connectors on the discharge and suction side 4 With types with coarse fine bleed valve Firstly remove the coarse fine bleed valve grip then lift off the cover of the liquid end with a screw driver P_G...

Page 71: ...direction until it is sitting tightly 17 Place the dosing head 2 with the screws 1 onto the diaphragm 3 and the backplate 4 ensure that the suction connector points downwards when the pump is in its...

Page 72: ...lphone material 4 Test the connected diaphragm rupture indicator fully immerse the tapered section at the front in water The continuous display indicates a diaphragm rupture 5 Thoroughly dry the diaph...

Page 73: ...ome into con tact with oxygen 15 1 Faults Without a Fault Message Fault description Cause Remedy Personnel Pump does not prime in spite of full stroke motion and bleeding Minor crystalline deposits on...

Page 74: ...nalogue menu and the control current has fallen below 4 mA Clear the cause of the low control current or switch the programming of the fault behaviour to off see chapter Set Up Settings Operating mode...

Page 75: ...m rupture The diaphragm is broken Replace the diaphragm and clean the diaphragm rupture indicator refer to the Repair chapter Technical per sonnel The identifier p appears followed by the message Over...

Page 76: ...e working on the pump WARNING Fire hazard with flammable media Only with combustible media These may start to burn when combined with oxygen During filling and draining of the liquid end an expert mus...

Page 77: ...e environment Separate the battery from the remaining parts Note the pertinent regulations currently applicable in your country CAUTION Environmental hazard due to electronic waste There are electroni...

Page 78: ...5 0414 4 12 0 1 00 8x5 2 0 1 8 0 5 0220 2 18 0 1 50 12x9 2 0 1 8 0 5 1009 10 8 0 0 67 8x5 3 0 1 8 0 5 0715 7 12 0 1 00 8x5 2 5 1 8 0 5 0424 4 20 0 1 67 12x9 2 5 1 8 0 5 gamma X metering pumps with do...

Page 79: ...outgassing media and when operating with air bubbles no dosing accuracy or reproducibility can be provided The recommended minimum stroke length with self bleeding dosing pumps is 50 17 3 Viscosity Th...

Page 80: ...04 Ceramic PTFE Ceramic Self bleeding design only in material versions PP and NP with a valve spring made of Hastelloy C and a PVDF valve insert Diaphragm with PTFE coating FKM fluorine rubber Housing...

Page 81: ...0068 2 30 2005 17 8 Degree of Protection and Safety Requirements Protection against accidental contact and humidity IP 65 according to DIN EN 60529 Degree of protection 1 mains power connection with p...

Page 82: ...to the gamma L product range is not guaranteed for the fol lowing points n External dimensions of the pump n Distance between the support surface for the mounting foot and the centre of the dosing he...

Page 83: ...18 Exploded Drawings and Ordering Information 18 1 Exploded drawings 1 5 1 8 6 P_G_0032_SW Fig 39 Liquid end gamma X 1602 1604 PP_2 Exploded Drawings and Ordering Information 83...

Page 84: ...PT2 Liquid end 1050931 1050900 1050942 Spare parts kit 1001646 1001654 1023109 Diaphragm 1000246 1000246 1000246 PPE2 PPB2 PPT2 Liquid end 1050932 1050901 1050943 Spare parts kit 1039989 1039987 10353...

Page 85: ...Spare parts for liquid end gamma X 0708 1009 0220 0424 PP_2 Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 0708 1009 0220 0424 PP_2 Exploded Drawings...

Page 86: ...50903 1050945 Spare parts kit 1001649 1001657 1023112 Diaphragm 1000249 1000249 1000249 PPE2 PPB2 PPT2 Liquid end 1050935 1050924 1050946 Spare parts kit 1051096 1051085 1051129 Diaphragm 1045456 1045...

Page 87: ...re parts for liquid end gamma X 0245 PP without bleed valve Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 0245 PP without bleed valve Exploded Drawi...

Page 88: ...re parts kit 1051097 1051086 1051130 Diaphragm 1045443 1045443 1045443 1 5 1 8 6 P_G_0035_SW Fig 42 Liquid end gamma X 0245 without bleed valve Liquid end gamma X 1602 2504 NP with and without bleed v...

Page 89: ...001723 1023109 Diaphragm 1000246 1000246 1000246 NPE NPB NPT Liquid end with bleed valve 1051175 1051163 1051179 Liquid end without bleed valve 1051176 1051174 1051180 Spare parts kit 1039988 1039986...

Page 90: ...gamma X 0708 1009 0220 0424 NP with and without bleed valve Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 0708 1009 0220 0424 NP with and without bl...

Page 91: ...0957 1050991 Spare parts kit 1001718 1001726 1023112 Diaphragm 1000249 1000249 1000249 NPE NPB NPT Liquid end with bleed valve 1050969 1050952 1050986 Liquid end without bleed valve 1050975 1050958 10...

Page 92: ...for liquid end gamma X 0245 NP with and without bleed valve Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 0245 NP with and without bleed valve Explo...

Page 93: ...0 1050953 1050987 Liquid end without bleed valve 1050976 1050959 1050993 Spare parts kit 1051119 1051108 1051130 Diaphragm 1045443 1045443 1045443 Liquid end gamma X 0245 with and without bleed valve...

Page 94: ...parts for liquid end gamma X 1602 1604 PVT with bleed valve Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 1602 1604 PVT with bleed valve Exploded Dr...

Page 95: ...parts kit 1023109 Diaphragm 1000246 PVT Liquid end 1050995 Spare parts kit 1035332 Diaphragm 1034612 Liquid end gamma X 1602 with bleed valve Liquid end gamma X 1604 with bleed valve Exploded Drawings...

Page 96: ...r liquid end gamma X 0708 1009 0220 0424 PVT with bleed valve Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 0708 1009 0220 0424 PVT with bleed valve...

Page 97: ...pare parts kit 1023112 Diaphragm 1000249 PVT Liquid end 1050998 Spare parts kit 1051129 Diaphragm 1045456 Liquid end gamma X 0708 1009 with bleed valve Liquid end gamma X 0414 0715 with bleed valve Li...

Page 98: ...parts kit for liquid end gamma X 0245 PVT without bleed valve Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 0245 PVT without bleed valve Exploded Dr...

Page 99: ...s kit 1051130 Diaphragm 1045443 1 1 6 P_G_0049_SW P_G_0049_SW Fig 48 Liquid end gamma X 0245 without bleed valve Liquid end gamma X 1604 0220 0424 PV HV for high viscosity feed chemicals Exploded Draw...

Page 100: ...s kit 1035342 Diaphragm 1034612 PV HV Liquid end 1051001 Spare parts kit 1019067 Diaphragm 1000248 PV HV Liquid end 1051002 Spare parts kit 1019069 Diaphragm 1000249 PV HV Liquid end 1051003 Spare par...

Page 101: ..._SW Fig 49 Spare parts for liquid end gamma X 1602 1604 TTT Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 1602 1604 TTT Exploded Drawings and Orderi...

Page 102: ...nd 1051016 Spare parts kit 1001739 Diaphragm 1000246 TTT Liquid end 1051017 Spare parts kit 1035331 Diaphragm 1034612 Liquid end gamma X 1602 Liquid end gamma X 1604 Exploded Drawings and Ordering Inf...

Page 103: ...W Fig 50 Liquid end gamma X 0708 1009 0220 0424 TTT Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 0708 1009 0220 0424 TTT Exploded Drawings and Orde...

Page 104: ...TTT Liquid end 1051019 Spare parts kit 1001742 Diaphragm 1000249 TTT Liquid end 1051020 Spare parts kit 1051151 Diaphragm 1045456 Liquid end gamma X 0708 1009 Liquid end gamma X 0414 0715 Liquid end...

Page 105: ...043_SW Fig 51 Spare parts kit for liquid end gamma X 0245 TTT Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 0245 TTT Exploded Drawings and Ordering...

Page 106: ...TTT Liquid end 1051021 Spare parts kit 1051152 Diaphragm 1045443 1 8 6 5 1 P_G_0044_SW Fig 52 Liquid end gamma X 0245 Liquid end gamma X 1602 2504 SST Exploded Drawings and Ordering Information 106...

Page 107: ...5 Discharge valve 6 Diaphragm 8 Suction valve SST Liquid end 1051004 Spare parts kit 1001731 Diaphragm 1000246 SST Liquid end 1051005 Spare parts kit 1035331 Diaphragm 1034612 Liquid end gamma X 1602...

Page 108: ...pare parts kit for liquid end gamma X 0708 1009 0220 0424 SST Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 0708 1009 0220 0424 SST Exploded Drawing...

Page 109: ...SST Liquid end 1051007 Spare parts kit 1001734 Diaphragm 1000249 SST Liquid end 1051008 Spare parts kit 1051139 Diaphragm 1045456 Liquid end gamma X 0708 1009 Liquid end gamma X 0414 0715 Liquid end...

Page 110: ...046_SW Fig 54 Spare parts kit for liquid end gamma X 0245 SST Item Description 1 Connector kit 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 0245 SST Exploded Drawings and Ordering...

Page 111: ...09 Spare parts kit 1051140 Diaphragm 1045443 1 4 5 8 10 9 6 P_G_0047_SW Fig 55 Liquid end gamma X 0245 Liquid end gamma X 1602 1604 PP SEK and 1602 1604 NP SEK self bleeding Exploded Drawings and Orde...

Page 112: ...1604 1050978 NPB SEK 1602 1050960 1604 1050961 Spare parts kits for type Material version Order no 1602 with spring PPE 1001757 1604 with spring PPE 1035335 1602 with spring PPB 1001763 1604 with spri...

Page 113: ..._0048_SW Fig 56 Item Part 1 9 10 Connector kit 4 Bleed valve 5 Discharge valve 6 Diaphragm 8 Suction valve Liquid end gamma X 0708 1008 0220 0420 PP NP SEK self bleeding Exploded Drawings and Ordering...

Page 114: ...050981 NPB SEK 0708 1009 1050962 0414 0715 1050963 0220 0424 1050964 Spare parts kits for type Material version Order no 0708 1008 PPE 1001759 0413 0713 PPE 1001760 0220 0420 PPE 1051102 0708 1008 PPB...

Page 115: ...dicating relay 1050643 Fault indicating pacing relay 1050654 Further information on spare parts accessories and options can be found in n the exploded drawings n the identity code n under www prominen...

Page 116: ...ings Compare the dimensions on the dimensional drawing with those of the pump and mounting foot All dimensions are in mm P_G_0065_SW 40082099 Fig 57 Dimensional drawing of gamma X mounting foot Dimens...

Page 117: ...04 0715 0414 0245 A 70 90 90 70 90 110 B 71 74 76 71 74 76 C with bleed valve 106 108 110 106 107 C without bleed valve 93 D 32 24 24 32 24 14 E 23 13 15 13 15 5 F 198 202 202 198 202 209 Dimensional...

Page 118: ...04 0715 0414 0245 A 70 90 90 70 90 110 B 77 74 76 77 76 76 C with bleed valve 105 102 104 105 104 105 C without bleed valve 92 91 91 92 91 91 D 33 23 23 33 23 14 E 191 200 200 191 200 210 Dimensional...

Page 119: ...9 0708 0424 220 1604 0715 0414 0245 A 70 90 90 70 90 110 B 71 75 79 71 73 76 C with bleed valve 84 92 90 84 90 C without bleed valve 93 D 36 25 25 36 25 14 E 196 203 203 196 203 209 Dimensional drawin...

Page 120: ...Fig 61 gamma X M70 gamma X M85 1009 0708 0424 220 1604 0715 0414 A 80 85 70 80 B 78 79 76 78 C 96 97 94 96 D 33 30 38 33 E 190 193 185 190 Dimensional drawing of gamma X mate rial version PVT HV Dime...

Page 121: ...0 gamma X M85 1602 1009 0708 0424 220 1604 0715 0414 0245 A 70 85 85 70 85 100 B 72 79 77 72 77 78 C 86 96 96 86 96 97 D 35 3 3 35 3 5 E 188 220 202 188 220 227 Dimensional drawing of gamma X mate ria...

Page 122: ...gamma X M85 1602 1009 0708 0424 220 1604 0715 0414 0245 A 70 85 85 70 85 100 B 72 79 77 72 77 78 C 84 94 94 84 94 95 D 42 11 11 42 11 4 E 180 211 211 180 211 219 Dimensional drawing of gamma X mate ri...

Page 123: ...ma X M70 gamma X M85 1602 1009 0708 0424 220 1604 0715 0414 A 70 90 90 70 90 B 77 74 76 77 76 C 92 89 91 92 91 D 33 24 24 33 24 E 189 199 199 189 199 Dimensional drawing of gamma X mate rial version P...

Page 124: ...ma X M70 gamma X M85 1602 1009 0708 0424 220 1604 0715 0414 A 70 90 90 70 90 B 77 76 76 77 76 C 92 91 91 92 91 D 33 24 24 33 24 E 189 199 199 189 199 Dimensional drawing of gamma X mate rial version N...

Page 125: ...1 min C l h f s GMXa 0708 C l h f s GMXa 0414 C l h f s GMXa 1602 200 1 min 180 1 min 160 1 min 140 1 min 120 1 min 100 1 min 80 1 min 60 1 min 40 1 min 20 1 min 200 1 min 180 1 min 160 1 min 140 1 m...

Page 126: ...n 60 1 min 40 1 min 20 1 min 200 1 min 180 1 min 160 1 min 140 1 min 120 1 min 100 1 min 80 1 min 60 1 min 40 1 min 20 1 min 200 1 min 180 1 min 160 1 min 140 1 min 120 1 min 100 1 min 80 1 min 60 1 m...

Page 127: ...in 60 1 min 40 1 min 20 1 min 0 5 10 15 20 25 0 10 20 30 40 50 60 70 80 90 100 90 80 70 100 50 40 30 20 10 60 C l h f s GMXa 0424 200 1 min 180 1 min 160 1 min 140 1 min 120 1 min 100 1 min 80 1 min 6...

Page 128: ...will invalidate this dec laration Extract from the EC Declaration of Conformity Designation of the product Metering pump Gamma X product range Product type GMXa _ _ _ _ _ _ _ _ _ _ _ _ U _ _ Y _ 0 Y 0...

Page 129: ...t pump Prime Start batch only in Batch operating mode Acknowledge errors Check adjustable values 2 Change directly changeable variables Settings operating mode Automatic Stroke length Metering Calibra...

Page 130: ...Service Language B0606 Operating Set up Overview of the gamma X 130...

Page 131: ...lume Total number of strokes Current stroke volume Settings Operating mode Manual Contact Memory On Off Metering volume 1 Factor 2 Batch Memory On Off Metering volume 1 Factor 2 Analogue 0 20 mA 4 20...

Page 132: ...alibration Calibration ended Calibration result System Dosing head No dosing head 0245 Volume unit Litres Gallons Pressure unit bar psi Inputs outputs Auxiliary capacity 1 Auxiliary frequency 2 Relay...

Page 133: ...Activation if auxiliary Diaphragm rupture Inactive Warning Error Pause Input normally open NO normally closed NC Level Warning normally open NO normally closed NC Level Error normally open NO normally...

Page 134: ...unter Stroke counter Volume counter All Diaphragm replace ment Back To change position Display Brightness Contrast Factory settings Password Yes No Language English German French x00e7 Spanish x00f1 1...

Page 135: ...e Analogue Analogue Stroke rate h Stroke rate min Stroke length Capacity Starting batch Factor Contact volume Batch dosing time Time 12000 h 200 min 50 0 12 00 l h 16 12 21 16 12 21 5 1 250 L 12000 h...

Page 136: ...2 L 16 12 21 2015 03 27 16 12 21 2015 03 27 16 12 21 2015 03 27 16 12 21 2015 03 27 slow slow slow Auxiliary display Stroke rate Capacity Factor Stroke length Signal current at the input Dosing mode P...

Page 137: ...ee of protection 81 Demonstration purposes 56 Designation of the product 128 DFMa 51 Diagrams for Setting the Capacity 125 Diaphragm replacement 63 70 Diaphragm replacement position 63 Diaphragm ruptu...

Page 138: ...ming time 61 Protection against contact and humidity 81 Protection class 81 Protection menu 62 Pulse control 47 Pump active 58 Q Qualification of personnel 11 R Relay 23 38 57 Relay polarity 58 Relay...

Page 139: ...de band 50 V Viscosity 50 52 79 Volume Stroke 45 Volume counter 63 Volume unit 56 W Warning 58 Warning alert 16 38 Warning indicator 15 Warning indicator yellow 24 Warning messages 75 Warning sign 8 I...

Page 140: ...984586 2 en_GB 2015 ProMinent GmbH Im Schuhmachergewann 5 11 69123 Heidelberg Germany Germany Telephone 49 6221 842 0 Fax 49 6221 842 612 Email info prominent com Internet www prominent com...

Reviews: