The confirmation of the zero
correction
[0.0.0] alternates
with
[CAL].
7.
Remove the cuvette from the
sample chamber and empty the
cuvette completely
8.
Thoroughly clean the cuvette
and the cuvette lid
9.
Fill the cuvette up to the 10-ml
marking with a standard solution
of known concentration
10. Close the cuvette using the cuv‐
ette lid
11. Position the cuvette in the
sample chamber
ð
Observe the correct posi‐
tioning of the cuvette.
12. Press the key
[ZERO/TEST]
ð
[METHOD] flashes for
approximately 3 seconds.
The confirmation of the
result alternates with
[CAL].
13. If the result matches the value
of the standard used, (within the
tolerance under consideration)
you can quit calibration mode by
pressing the
[ON/OFF] key
Changing the displayed value:
Pressing the [MODE] key 1 x
increases the displayed result by
1 digit.
Pressing the [ZERO/TEST] key 1
x reduces the displayed result by
1 digit.
1.
Repeatedly press the keys until
the displayed result matches the
value of the standard used
2.
Press the key
[ON/OFF]
ð
The new correction factor is
calculated and saved in the
user calibration level.
The confirmation of the cali‐
bration appears in the dis‐
play for 3 seconds.