5
Calibration
1.
The device is switched off.
Press and hold down the
[MODE] key
2.
Use the
[ON/OFF] key to switch
on the device
ð
3 decimal points appear in
the display.
3.
Release the
[MODE] key
4.
The [!] key allows you to select
the following operating menu
items:
n
[diS] = read out of stored
data
n
Set date and time
n
User calibration
ð
The selected menu item is
indicated by an arrow in the
display.
5.
Use the
[!] key to select the
[CAL] operating menu item
(arrow top right in the display)
User calibration
NOTICE!
A separate calibration of the
chlorine dioxide or chlorite meas‐
uring ranges is not possible. The
chlorine (CL) measuring range
calibration is relied on.
User calibration (display in calibration
mode ) =
[cCAL]
Factory calibration (display in calibra‐
tion mode ) =
[cCAL]
1.
Confirm the selection by
pressing the
[MODE] key
ð
The display toggles between
[CAL / METHOD]
2.
Select the method to be cali‐
brated using the
[MODE] key
3.
Fill the cuvette up to the 10 ml
marking with the standard solu‐
tion
4.
Close the cuvette using the cuv‐
ette lid
5.
Position the cuvette in the
sample chamber
ð
Observe the correct posi‐
tioning of the cuvette
6.
Press the key
[ZERO/TEST]
ð
[METHOD] flashes for
approximately 8 seconds.
Calibration
31