13. Hold the dropping bottle upright
and by slow pressing, add equal
sized drops into the cuvette:
n
6 drops ➨ DPD 1 buffer sol‐
ution
n
2 drops ➨ DPD 1 reagent
solution
Fill the cuvette up to the 10 ml
marking with the water sample
14. Close the cuvette using the cuv‐
ette lid
15. Mix the contents of the cuvette
by tipping it back and forth
16. Position the cuvette in the
sample chamber
ð
Observe the correct posi‐
tioning of the cuvette
17. Press the key
[ZERO/TEST]
ð
[METHOD] flashes for
approximately 3 seconds.
The display value
[A] = sum
of free chlorine plus chlorine
dioxide appears in the dis‐
play.
18. Remove the cuvette from the
sample chamber
19. Hold the dropping bottle upright
and by slow pressing, add equal
sized drops into the cuvette:
n
3 drops ➨ DPD 3 solution
20. Close the cuvette using the cuv‐
ette lid
21. Mix the contents of the cuvette
by tipping it back and forth
22. Position the cuvette in the
sample chamber
ð
Observe the correct posi‐
tioning of the cuvette
23. Switch the countdown on, see
function on’ on page 13. To do
this, press the
[!] and
[ZERO/TEST] key
ð
Wait for the 2 minutes reac‐
tion time to elapse
24.
[METHOD] flashes for approxi‐
mately 3 seconds.
ð
The display value
[C] = sum
of free chlorine plus com‐
bined chlorine plus chlorine
dioxide appears in the dis‐
play.
25. Calculation:
n
Chlorine dioxide (mg/l) =
display value
[G] x 1.9
n
Free chlorine (mg/l) = dis‐
play value
[A] minus display
value
[G]
n
Combined chlorine (mg/l) =
display value
[C] minus dis‐
play value
[A]
Analysis methods
21