ProMinent DULCOMETER D1C A D Series Operating Instructions Manual Download Page 23

23

Complete Operating Menu / Description

Possible values

Initial value

Increment

Lower value

    Upper value

Remarks

Alarm relay

active

active

not active

Pause

normal

normal

Hold

Control input pause

active closed

active closed

Reacts as make contact

active open

Reacts as break contact

Pause with alarm

off

off

Alarm relay can be

on

triggered by pause contact

td

0 min

1 min

0 min

60 min

Access code

5000

1

1

9999

Language

as per identity

as per identity

code

code

Operating menu

complete

restricted

complete

Standard Pause

If the pause switch is off, the DULCOMETER

®

 D1C sets the operating outputs to “0” for as long as the

pause switch is off or for a set time-delay t

d

 (if t

d

 is set to > 0 min). Whilst the pause switch is off, the D1C

establishes the P-proportion in the background.
With PID control (Identity code characteristics “control characteristic” = 2): the I-proportion is stored when
the pause is switched off (I-proportion then usually only present if Ti > 0 has been selected in the “Control
setting?” setting menu).
Exception: the standard signal outputs mA for the measured value or correction value are not affected by
the pause.
After pause is activated, the operating outputs remain at “0” for the length of the time delay t

d

. The time

delay t

d

  must be set up in such a way that in this time e.g. sample water (process-specific current

concentration) flows to the sensor.
With PID control (Identity code characteristics “control characteristic” = 2): The control variable output
resulting from the pause and the expiry of the time-delay t

d

  is reconciled jointly with the current

P-component and (if Ti is set > 0) with the stored I-component.

Pause Hold

If the pause switch is off, the DULCOMETER

®

 D1C freezes the operating output at the most recent value for

as long as the pause switch is off or for a set time-delay t

d

 (if t

d

 is set to > 0 min). Whilst the pause switch is

off, the D1C establishes the P-proportion in the background.
With PID control (Identity code characteristics “control characteristic” = 2):
Even the mA standard signal outputs for measured value or correction value are frozen.
After pause is activated, the operating outputs remain frozen for the length of the time delay t

d

. The time

delay t

d

  must be set up in such a way that in this time e.g. sample water (process-specific current

concentration) flows to the sensor.
With PID control (Identity code characteristics “control characteristic” = 2): The control variable output
resulting from the pause and the expiry of the time-delay t

d

  is reconciled jointly with the current

P-proportion and (if Ti is set > 0) with the newly established I-proportion.

Access Code

Access to the setting menu can be prevented by setting up an access code. The DULCOMETER

®

 D1C

controller is supplied with the access code 5000 which permits free access to the setting menu. The
calibration menu remains freely accessible even if access to the setting menu is blocked by the code.

BA_DM_183_10_10_EN.p65

28.10.2010, 11:44 Uhr

23

Summary of Contents for DULCOMETER D1C A D Series

Page 1: ...850 ProMinent Dosiertechnik GmbH 69123 Heidelberg Germany BA DM 183 10 10 EN Please completely read through operating instructions Do not discard The operator shall be liable for any damage caused by...

Page 2: ...mV Correction variable 0 None Feed forward control 0 None 1 via standard signal 0 4 20 mA 2 via frequency 0 500 Hz 3 via frequency 0 10 Hz Control input 0 None 1 Pause Signal output 0 None 1 standard...

Page 3: ...data and function of the series DULCOMETER D1C controller provide detailed safety information and are divided into clear steps IMPORTANT Please observe the parts of these operating instructions appli...

Page 4: ...button To change over within a menu level and to change from one variable to another within a menu point START STOP button Start stop of control and metering function ENTER button To accept confirm or...

Page 5: ...riable starts again after cessation of pause 4 4 Feed Forward Control The D1C controller can process a signal of a feed forward control Depending on the device version see identity code and the settin...

Page 6: ...off left Symbol Control on left Metering pump 2 reducing agent Symbol Control off right Symbol Control on right Solenoid valve 1 oxidant Symbol Control off left Symbol Control on left Solenoid valve...

Page 7: ...Operation Access code correct Parameter setting Calibration notes Permanent display 1 Permanent display 2 Calibration menu Various Access code Setting menus D1C2 007 GB The various menus are selected...

Page 8: ...er Ti Td xp 0 s 750 mV setpoint 1 lower 650 mV 10 ctrl parameter xp 10 control settings Only with control control positive oxide regulated value negative reduc manual dosing control current regulat va...

Page 9: ...ys After everything has been checked all error tests which refer to the measured value are restarted Restricted Operating Menu Description Error message Condition Effect Measured value high Measured v...

Page 10: ...value 30 control current regulat value 30 D1C2 015 Redox GB Access to all setting menus can be blocked with an access code Only with control Note The controlled variable is recalculated every second...

Page 11: ...ode 5000 1 1 9999 Language as per identity as per identity code code Operating menu restricted restricted complete D1C2 016 GB Access to all setting menus can be blocked with an access code general se...

Page 12: ...Only with servomotor Only with pumps meas value setting pumps setting relays setting limits setting servomotor setting Number and scope of setting menus is dependent on the device Access to setting me...

Page 13: ...V 2000 mV if measured variables 2 and 5 185 265 mV 220 mV are connected 425 505 mV 465 mV 1000 mV if measured variable 1 is connected Complete Operating Menu Description D1C2 018 Redox GB Permanent di...

Page 14: ...ng value input changes within the control time measuring value It is assumed that the value does not change for an intact sensor If the measuring value does not change during this control time the DUL...

Page 15: ...ump1 pump2 control ORP pump1 oxide pump2 reduc pulse minute 180 180 In the setting menu correction value for this equipment enables you to display the temperature or to maintain an mA signal proprotio...

Page 16: ...motor off Relay 2 Solenoid valve 2 Limit value 2 Actuator 2 Timer 2 off Cycle 10 s 1 s 10 s 9999 s for solenoid valve min time 1 s 1 s 1 s Cycle 2 for solenoid valve Set here the smallest permitted op...

Page 17: ...n time smallest permitted operating factor of the connected device The actuating variable determines the ratio ton cycle and thus the switching times see fig above The min time influences the switchin...

Page 18: ...Actuating direction active closed active closed Reacts as make contact limit value 1 active open Reacts as break contact limit value 2 Switch on delay t on 0 s 1 s 0 s 9999 s Switch off delay t off 0...

Page 19: ...er value Remarks Servomotor Setting Setting ok off Control direction Oxidant Oxidant Reducing Control range 100 1 10 100 in of operating range D1C2 025 Redox GB Access to all setting menus can be bloc...

Page 20: ...026 Redox GB 10 control load regulated range Access to all setting menus can be blocked with an access code Only with control Note The controlled variable is recalculated every second Only suitable fo...

Page 21: ...ltiplicative multiplicative Feed forward control effect additive Max add regulated only with add feed variable 100 1 500 500 forward control D1C2 027 Redox GB Access to all setting menus can be blocke...

Page 22: ...tting menus can be blocked with an access code general setting information ident code D1CA software version DxRxxxxxxxxxx D1C A1 FW 5 00 alarm relay access c operating menu active 5000 english complet...

Page 23: ...ows to the sensor With PID control Identity code characteristics control characteristic 2 The control variable output resulting from the pause and the expiry of the time delay td is reconciled jointly...

Page 24: ...error mV calibration Basic load Stop No Metering continues in Check probe replace if faulty case of error with unsteady necessary recalibrate measured values if necessary Correction measured variable...

Reviews: