18
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Type of limit transgression
upper
Limit transgression when
Limit 1:
lower
lower
exceeding or dropping below value
Limit 2:
upper
off*
*only with limit relay
Limit value
Limit 1:
500 mV
1 mV
-2000 mV
2000 mV
Limit 2:
1000 mV
1 mV
-2000 mV
2000 mV
Hysteresis limits
10 mV
1 mV
10 mV
2000 mV
Effective in direction of cancelling
limit transgression
Checkout time limits
off
1 s
1 s
9999 s
Results in message and alarm.
off = 0 s:
Function switched off,
no message, no alarm
Control
on
on
off
Actuating direction
active closed
active closed
Reacts as make contact
limit value 1;
active open
Reacts as break contact
limit value 2
Switch-on delay
∆
t on
0 s
1 s
0 s
9999 s
Switch-off delay
∆
t off
0 s
1 s
0 s
9999 s
Complete Operating Menu / Description
Limits
D1C2-024-Redox-GB
relay 2
LV2
- active
closed
∆
t on
0 s
∆
t off
0 s
Access to all setting menus can be blocked with an access code !
limits
setting ?
relay 1
LV 1
- active closed
∆
t on
0 s
∆
t off
0 s
limit 2
upper
800 mV
limit 1
lower
100 mV
hysteresis
limits:
10 mV
∆
t on:
off
control: on
If the limit is exceeded for longer than the “Delay time - limit values”, an error message is given which
must be acknowledged, and the alarm relay circuit is broken. If “Control” is also set to “off” the control
process stops.