22
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Variable allocation
as per identity
Measured value
code
Controlled variable
If control applicable
Output range
0…20 mA
0…20 mA
4…20 mA
3.6/4-20 mA
Reduction to 3.6 mA
when alarm relay switches
(not limit value violation)
Range measured
value
0–1 V
1 mV
-2000 mV
2000 mV
Minimum range 10 mV
Range controlled
variable
0 %…+100 %
1 %
-100 %
+100 %
Minimum range 1 % of
measured value
General Setting
D1C2-028-GB
Access to all setting menus can be blocked with an access code !
general setting
information
ident-code: D1CA
software version
DxRxxxxxxxxxx
D1C-A1 FW-5.00
alarm relay
access c.:
=
=
operating menu:
active
5000
english
complete
pause normal
- active... closed
- alarm off
- td: 0 min
Standard Signal Output 2
D1C2-030-Redox-GB
Access to all setting menus can be blocked with an access code !
Control with standard signal
signal output 2
mA setting ?
Only with 2 standard signal outputs
signal output 2
mA
0…20 mA
control
positive oxide
regulated value:
negative reduc
signal output 2
0%
0 mA =
20 mA =
regulated value
100%
measured value
signal output 2
1000 mV
0 mA =
20 mA =
measured value
0 mV
Complete Operating Menu / Description
BA_DM_183_10_10_EN.p65
28.10.2010, 11:44 Uhr
22