Troubleshooting 

For questions not addressed here, please contact your local Promega Branch Office or Distributor. 
Contact information available at: 

www.promega.com

. E-mail: 

[email protected] 

5.A. Common Problems 

Symptoms

Causes and Comments

 

Low or no fluorescence reading

Reagents were not prepared correctly. Ensure 
that reagents, blank and standard samples are 
prepared as described in Section 3.A. Check 
that the reagents have not expired. Check that 
the reagents and standard were stored correctly. 

Fluorometer was not calibrated correctly. Be 
sure to calibrate any protocol being used for the 
first time (see Sections 3.A and 3.C). We 
recommend recalibrating the instrument with 
each new lot of reagent. Calibration with each 
use will reduce experimental variation (e.g., due 
to pipetting). 

The wrong standard concentration was used to 
calibrate the Quantus™ Fluorometer. Check 
that the fluorescence is within the range of the 
standard curve used (e.g., if the unknown 
samples are of low concentration, then use the 
low concentration standard), and that the 
standard used is similar to the sample (i.e., if 
the unknown sample is plasmid DNA, then use 
a plasmid DNA standard). 

Sample concentration was too low. Redilute 
sample, and repeat measurement. 

A PCR tube with poor optical clarity was used.  
Thin-walled PCR tubes have the best optical  
clarity. Ensure a thin-walled PCR tube was  
used. We recommend using thin-walled 0.5ml  
PCR tubes (Cat.# E4942 or Axygen  
Cat.# PCR-05-C, available from Fisher or VWR)  
for proper optical clarity. 

Quantus™ Fluorometer will not

Instrument was not plugged in. Make sure that 

power on

the instrument is plugged into the USB Power 
Supply or a computer that is turned on. 

Use an alternative power source. If the  
fluorometer was plugged into the USB power  
supply, switch to a computer or vice versa. If  
switching power sources does not power on the  
Quantus™ Fluorometer, contact Promega  
Technical Services. 

 

Promega Corporation

 

·

 2800 Woods Hollow Road 

·

 Madison, WI 53711-5399 USA  

Toll Free in USA 800-356-9526 

·

 Phone 608-274-4330 

·

 Fax 608-277-2516 

·

 

www.promega.com

 

Printed in USA.

Part# TM396 

Revised 1/20

Page 15

TM396.0120_EIVD_TM.qxd  1/27/2020  2:50 PM  Page 15

Summary of Contents for E6150

Page 1: ...T e c h n i c a l M a n u a l Quantus Fluorometer Operating Manual INSTRUCTIONS FOR USE OF PRODUCT E6150 PRINTED IN USA Revised 1 20 Part TM396 TM396 0120_EIVD_TM qxd 1 27 2020 2 50 PM Page a...

Page 2: ...shooting 15 A Common Problems 15 B Error Messages 16 6 Appendix 17 A Updating Firmware 17 B Warranty and Service 17 C Certificate of Decontamination 18 D Related Products 19 7 Summary of Change 19 1 I...

Page 3: ...y Includes Quantus Fluorometer USB Power Supply USB Cord Outlet Adapter Kit 4 adapters supplied for international compatibility Quick Reference Guide Operating Manual Product Registration Card Shippin...

Page 4: ...on sensor Excitation Peak Red 625nm Blue 470nm Excitation Filters Red 640nm shortpass wavelengths up to 640nm Blue 495nm shortpass wavelengths up to 495nm Emission Filters Red 660 720nm Blue 510 580nm...

Page 5: ...ques Veuillez contacter votre repr sentant Promega local concernant la destruction de l appareil Veuillez respecter les exigences de votre tablissement concernant la destruction des accessoires 9630TA...

Page 6: ...nt may be impaired Do not disassemble the unit The Quantus Fluorometer accepts 0 5ml PCR tubes Do not force larger tubes into the tube holder Use caution around solvents because they may corrode the p...

Page 7: ...etting Up the Quantus Fluorometer 1 Position the Quantus Fluorometer on a flat surface 2 Select the appropriate outlet adapter for your wall outlet from the Outlet Adapter Kit provided Promega Corpora...

Page 8: ...un setup exe Follow prompts to install the software If NET 4 0 framework is not installed on the computer the software installer will prompt the user to download and install the framework software ava...

Page 9: ...s Fluorometer protocol has calibration values associated with the specific assay reagent Once the protocol of interest is calibrated the Quantus Fluorometer will save the calibration values for future...

Page 10: ...0 single concentration standard curve 1 400 RNA Cat E3310 high concentration standard curve 1 400 RNA Cat E3310 low concentration standard curve 1 2 000 ssDNA Cat E3190 high concentration standard cur...

Page 11: ...electing the desired protocol navigate to the Calibration screen 2 Place the blank sample into the tube holder and close the lid Select Read Blank and the fluorescence in relative fluorescence units R...

Page 12: ...tered is 100 200 0 5 4 Choose the appropriate fluorescence channel Red Excitation 640nm shortpass emission 660 720nm Blue Excitation 495nm shortpass emission 510 580nm 5 Select Save Note To edit or de...

Page 13: ...it is within the linear range of the standard 1 Press any button to wake up the Quantus Fluorometer The Home screen will appear displaying the last protocol used the last concentration measurement an...

Page 14: ...esult history select Clear from the menu list Results can be transferred to a connected computer as the measurements are read using the Quantus Software see Section 2 B The default fields transferred...

Page 15: ...f additional cleaning is needed wipe with a cloth dampened with a mild detergent remove detergent with a cloth dampened with water and wipe dry 4 Do not use organic solvents they may damage the plasti...

Page 16: ...ow concentration then use the low concentration standard and that the standard used is similar to the sample i e if the unknown sample is plasmid DNA then use a plasmid DNA standard Sample concentrati...

Page 17: ...o retain the calibration values Valid on Calibration screen The standard to blank ratio was within the parameters expected by the instrument Press Save to retain the calibration values and proceed to...

Page 18: ...e update only when instructed by Promega Technical Services Follow the instructions located at www promega com quantusfluorometerfirmware 6 B Warranty and Service The Quantus Fluorometer comes with a...

Page 19: ...aminated and can be handled without exposing personnel to health hazards Circle the type of material used in the instrument Chemical Biological Radioactive Briefly describe the decontamination procedu...

Page 20: ...00 356 9526 Phone 608 274 4330 Fax 608 277 2516 www promega com Printed in USA Part TM396 Revised 1 20 Page 19 2013 2020 Promega Corporation All Rights Reserved QuantiFluor is a registered trademark o...

Page 21: ...Promega Corporation Promega Corporation 2800 2800 W Woods Hollow Road oods Hollow Road Madison WI 53711 5399 USA Phone 608 274 4330 TM396 0120_EIVD_TM qxd 1 27 2020 2 50 PM Page b...

Reviews: