background image

3.A. Calibrating the Quantus™ Fluorometer for Use With the QuantiFluor

®

 Dye 

Systems (continued) 

4. Prepare the blank sample for the QuantiFluor

®

 ONE dsDNA System by 

adding 200μl of QuantiFluor

®

 ONE dsDNA Dye to a 0.5ml PCR tube. 

Prepare the blank sample for all other QuantiFluor

®

 Systems by adding 

200μl of QuantiFluor

®

 Dye working solution prepared in Step 2 to a 0.5ml 

PCR tube. 

5. For the QuantiFluor

®

 ONE dsDNA System, add 1µl of the standard to 

200μl of QuantiFluor

®

 ONE dsDNA Dye. 

For all other QuantiFluor

®

 Systems, add standard prepared in Step 3 to 

200µl of QuantiFluor

®

 Dye working solution prepared in Step 2.  

6. Mix three times by pipetting slowly. When using aerosol-resistant pipette 

tips, do not allow the pipette tip filter to get wet. Alternatively, vortex 
tubes at a high setting for 10 seconds. 

7.

Optional:

 Centrifuge tubes at 2,000 × 

g

 for 5–10 seconds to collect liquid at 

the bottom of the tube and remove any bubbles present. 

8. Incubate tubes at room temperature for 5 minutes, protected from light. 

Calibration Protocol

 

1. Select the desired QuantiFluor

®

 Dye assay from the Protocol screen on the 

instrument. If this is the first time the protocol has been selected, the 
Calibration screen will automatically appear. Otherwise, after selecting the 
desired protocol, navigate to the Calibration screen. 

2. Place the blank sample into the tube holder, and close the lid. Select “Read 

Blank”, and the fluorescence in relative fluorescence units (RFU) for the 
blank sample will be displayed on the screen. 

3. Place the standard sample into the tube holder, and close the lid. Select 

“Read Std”, and the fluorescence in RFU for the standard sample will be 
displayed on the screen. 

Note:

 Once both the blank and standard samples have been measured, the 

standard-to-blank ratio is automatically calculated as a quality check to 
ensure the QuantiFluor

®

 Dye working solution and blank and standard 

samples were prepared properly. If the standard-to-blank ratio is lower 
than the minimal accepted ratio, the screen will display “Invalid”, and the 
instrument will not quantitate the unknown samples until the blank and 
standard samples are prepared properly. The minimum standard-to-blank 
ratio is: 

• 50 for high-concentration standard curve or single standard curve protocols 

• 4 for low-concentration standard curve protocols  

• 3 for user-defined protocols 
These ratios have been predetermined as a quality check to help ensure a 
successful measurement. 

Promega Corporation

 

·

 2800 Woods Hollow Road 

·

 Madison, WI 53711-5399 USA  

Toll Free in USA 800-356-9526 

·

 Phone 608-274-4330 

·

 Fax 608-277-2516 

·

 

www.promega.com

 

Part# TM396

Printed in USA. 

Page 10

Revised 1/20

TM396.0120_EIVD_TM.qxd  1/27/2020  2:50 PM  Page 10

Summary of Contents for E6150

Page 1: ...T e c h n i c a l M a n u a l Quantus Fluorometer Operating Manual INSTRUCTIONS FOR USE OF PRODUCT E6150 PRINTED IN USA Revised 1 20 Part TM396 TM396 0120_EIVD_TM qxd 1 27 2020 2 50 PM Page a...

Page 2: ...shooting 15 A Common Problems 15 B Error Messages 16 6 Appendix 17 A Updating Firmware 17 B Warranty and Service 17 C Certificate of Decontamination 18 D Related Products 19 7 Summary of Change 19 1 I...

Page 3: ...y Includes Quantus Fluorometer USB Power Supply USB Cord Outlet Adapter Kit 4 adapters supplied for international compatibility Quick Reference Guide Operating Manual Product Registration Card Shippin...

Page 4: ...on sensor Excitation Peak Red 625nm Blue 470nm Excitation Filters Red 640nm shortpass wavelengths up to 640nm Blue 495nm shortpass wavelengths up to 495nm Emission Filters Red 660 720nm Blue 510 580nm...

Page 5: ...ques Veuillez contacter votre repr sentant Promega local concernant la destruction de l appareil Veuillez respecter les exigences de votre tablissement concernant la destruction des accessoires 9630TA...

Page 6: ...nt may be impaired Do not disassemble the unit The Quantus Fluorometer accepts 0 5ml PCR tubes Do not force larger tubes into the tube holder Use caution around solvents because they may corrode the p...

Page 7: ...etting Up the Quantus Fluorometer 1 Position the Quantus Fluorometer on a flat surface 2 Select the appropriate outlet adapter for your wall outlet from the Outlet Adapter Kit provided Promega Corpora...

Page 8: ...un setup exe Follow prompts to install the software If NET 4 0 framework is not installed on the computer the software installer will prompt the user to download and install the framework software ava...

Page 9: ...s Fluorometer protocol has calibration values associated with the specific assay reagent Once the protocol of interest is calibrated the Quantus Fluorometer will save the calibration values for future...

Page 10: ...0 single concentration standard curve 1 400 RNA Cat E3310 high concentration standard curve 1 400 RNA Cat E3310 low concentration standard curve 1 2 000 ssDNA Cat E3190 high concentration standard cur...

Page 11: ...electing the desired protocol navigate to the Calibration screen 2 Place the blank sample into the tube holder and close the lid Select Read Blank and the fluorescence in relative fluorescence units R...

Page 12: ...tered is 100 200 0 5 4 Choose the appropriate fluorescence channel Red Excitation 640nm shortpass emission 660 720nm Blue Excitation 495nm shortpass emission 510 580nm 5 Select Save Note To edit or de...

Page 13: ...it is within the linear range of the standard 1 Press any button to wake up the Quantus Fluorometer The Home screen will appear displaying the last protocol used the last concentration measurement an...

Page 14: ...esult history select Clear from the menu list Results can be transferred to a connected computer as the measurements are read using the Quantus Software see Section 2 B The default fields transferred...

Page 15: ...f additional cleaning is needed wipe with a cloth dampened with a mild detergent remove detergent with a cloth dampened with water and wipe dry 4 Do not use organic solvents they may damage the plasti...

Page 16: ...ow concentration then use the low concentration standard and that the standard used is similar to the sample i e if the unknown sample is plasmid DNA then use a plasmid DNA standard Sample concentrati...

Page 17: ...o retain the calibration values Valid on Calibration screen The standard to blank ratio was within the parameters expected by the instrument Press Save to retain the calibration values and proceed to...

Page 18: ...e update only when instructed by Promega Technical Services Follow the instructions located at www promega com quantusfluorometerfirmware 6 B Warranty and Service The Quantus Fluorometer comes with a...

Page 19: ...aminated and can be handled without exposing personnel to health hazards Circle the type of material used in the instrument Chemical Biological Radioactive Briefly describe the decontamination procedu...

Page 20: ...00 356 9526 Phone 608 274 4330 Fax 608 277 2516 www promega com Printed in USA Part TM396 Revised 1 20 Page 19 2013 2020 Promega Corporation All Rights Reserved QuantiFluor is a registered trademark o...

Page 21: ...Promega Corporation Promega Corporation 2800 2800 W Woods Hollow Road oods Hollow Road Madison WI 53711 5399 USA Phone 608 274 4330 TM396 0120_EIVD_TM qxd 1 27 2020 2 50 PM Page b...

Reviews: