2. Prepare the QuantiFluor
®
Dye working solution with 1X TE buffer as
follows. For example, to make a 1:400 dilution, combine 10µl of
QuantiFluor
®
Dye with 3,990µl of 1X TE buffer, and mix.
3. Prepare the nucleic acid standard in a 0.5ml PCR tube. Use the volume of
supplied standard and volume of 1X TE buffer indicated in the table below
to prepare the standard.
Promega Corporation
·
2800 Woods Hollow Road
·
Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526
·
Phone 608-274-4330
·
Fax 608-277-2516
·
www.promega.com
Printed in USA.
Part# TM396
Revised 1/20
Page 9
QuantiFluor
®
Dye System
Dilution of QuantiFluor
®
Dye
ONE dsDNA (Cat.# E4871), single-
concentration standard curve
no dilution necessary
(dye is pre-diluted)
dsDNA (Cat.# E2670), single-
concentration standard curve
1:400
RNA (Cat.# E3310), high-
concentration standard curve
1:400
RNA (Cat.# E3310), low-
concentration standard curve
1:2,000
ssDNA (Cat.# E3190), high-
concentration standard curve
1:400
ssDNA (Cat.# E3190), low-
concentration standard curve
1:2,000
QuantiFluor
®
Dye System
Starting
Standard
Concentration
Volume of
Standard
Volume of
Working
Solution
Final
Standard
Calibration
(per tube)
ONE dsDNA (Cat.# E4871)
400ng/µl
1µl
200µl
400ng
dsDNA (Cat.# E2670)
100ng/µl
2µl
200µl
200ng
RNA (Cat.# E3310), high-
concentration standard curve
100ng/µl
5µl
200µl
500ng
RNA (Cat.# E3310), low-
concentration standard curve
100ng/µl
First dilute
1:100 in
1X TE
Buffer, then
use 10µl
200µl
10ng
ssDNA (Cat.# E3190), high-
concentration standard curve
100ng/µl
4µl
200µl
400ng
ssDNA (Cat.# E3190), low-
concentration standard curve
100ng/µl
First dilute
1:100 in
1X TE
Buffer, then
use 10µl
200µl
10ng