P
ROGRAM
S
ELECTION AND
S
TARTING THE
T
REATMENT
Programmable programs P-C1 and P-C2
1.
Press the “ON/OFF” button to switch on the unit.
2.
Keep pressing the “Program” button until the programmable
programs (P-C1 or P-C2) are shown on the display.
3.
Press the “Set” button once. In the display, the frequency (Hz)
and “SET” will start to flash. Use the “+” or “–“ buttons to enter
the desired value (1 Hz to 200 Hz).
4.
Press the “Set” button. In the display, the pulse width (µS) and
“SET” will start to flash. Use the “+” or “–“ buttons to enter the
desired value (from 30 µS to 400 µS) in steps of ten.
5.
Press the “Set” button once more. The time will flash on the dis-
play, “ ´ “ for minutes and “ `` “ for seconds. Use the “+” or
“–“ buttons to enter the desired treatment duration (1 minute to
24 hours).
6.
Press the “Set” button to confirm and then press the “Select”
button 2 times. The intensity display “CH1” will start to flash.
Use the “+” or “–“ buttons to enter the desired value (from 1 to
50). Press the “Select” button once more. The intensity display
“CH2” will start to flash. Once again, use the “+” or “–“ buttons
to set the desired value. Fast adjustment protection will be activa-
ted and stops at intensity level 15 when the intensity increases
rapidly.
7.
To begin treatment, press the “Select” button once more. The
clock symbol in the top right of the display will start to flash.
8.
If you want to change the intensity during treatment, press the
“Select” button again and adjust the intensity as required using
the “+” or “–“ buttons.
62