Problem
Possible Cause
Solution
It will not emit a light flash when
the release button is pressed.
The attachment is not properly
placed on the appliance.
Ensure that the attachment has been mounted
correctly.
The appliance is not yet
operational.
Makes sure that the indicator lamp START flashes
green.
Poor contact with the skin.
Ensure that the skin support contact fully rests on
the skin.
The lamp in the attachment is
defective.
Replace the attachment with a new one.
If the error persists, contact our customer service.
Smoke can be seen or smelled
during operation.
The appliance is faulty.
Immediately stop the usage!
Contact our service centre or a specialist.
The indicator lamp START
and one of the level displays
remain lit.
System error
Switch the appliance off. Disconnect the power
adapter from the socket for around 5 seconds. Re-
insert the power adapter into the socket and switch
the appliance back on.
If the error persists, contact our customer service.
The indicator lamp START is lit.
And the other indicator lamps
start flashing.
The appliance has reached the
end of its life.
Dispose the appliance properly. See chapter
“Disposal”.
Technical Data
Model:.................................................................PC-IPL 3024
Power supply:...................................................DC 12 V, 2,5 A
Light surface: ............................................................... 2.7 cm
2
Technology used: ..........................IPL (Intense Pulsed Light)
Flash duration: ..................................................500 to 800 µs
Intensity density: .........................................................4J / cm
2
Wavelength: ...................................................475 to 1200 nm
Net weight: ....................................................... approx. 0.4 kg
Temperature:
Operation: ................................................+10 °C to +35 °C
Storage: ................................................... -40 °C to +70 °C
Relative air humidity:
Operation: ....................................................3 % to 75 % rH
Storage: .....................................................10 % to 90 % rH
Atmospheric pressure:
Operation: .................................................700 to 1060 hPa
Storage: ....................................................500 to 1060 hPa
External Power Adapter
Input: ............................
AC 100 − 240 V, 50 / 60 Hz, 0,8 A Max
Output:.................................................................DC 12 V / 3 A
Polarity: ......................................................................
Protection class:: ...................................................................