PRODIGIT
3260A Series Operation Manual 45
BANK
Purpose
:
:
:
:
To set waveform bank.
Command syntax
:
:
:
:
BANK{SP}{d}{
;
|NL} d
:
0
~
10
Description
:
:
:
:
This command is to set the waveform bank desired to be selected.
1. waveform bank 0
~
2 are sine wave.
2. waveform bank 3
~
8 are P.F..
3. waveform bank 9 are square wave.
4. waveform bank 10 is DC.
5. There are five (5) waveform information for each waveform bank, therefore, there
is a total 55 waveform information for eleven (11) waveform banks. Waveform
information is shown in Table 4-4.
For Example
:
:
:
:
a. BANK 1 ; To set waveform bank 1.
b. BANK 10 ; To set waveform bank 10.
Wave
From Bank
A
B
C
D
E
sine wave
0
√
2
2.0
2.5
3.0
3.5
1
1.5
1.6
1.7
1.8
1.9
2
3.0
3.1
3.2
3.3
3.4
C.F.= 2.0
3
P.F.=-0.85 P.F.=-0.80 P.F.=-0.75 P.F.=-0.70 P.F.=-0.65
C.F.= 2.5
4
P.F.=-0.70 P.F.=-0.65 P.F.=-0.60 P.F.=-0.50 P.F.=-0.40
C.F.= 3.5
5
P.F.=-0.50 P.F.=-0.45 P.F.=-0.40 P.F.=-0.35 P.F.=-0.30
C.F.= 2.0
6
P.F.= 0.85 P.F.= 0.80 P.F.= 0.75 P.F.= 0.70 P.F.= 0.65
C.F.= 2.5
7
P.F.= 0.70 P.F.= 0.65 P.F.= 0.60 P.F.= 0.50 P.F.= 0.40
C.F.= 3.5
8
P.F.= 0.50 P.F.= 0.45 P.F.= 0.40 P.F.= 0.35 P.F.= 0.30
square wave
9
1.0
1.1
1.2
1.3
1.4
DC
10
√
2dc
2dc
2.5dc
3.0dc
3.5dc
Table 4-4 waveform information
WAVE
Purpose
:
:
:
:
To set waveform.
Command syntax
:
:
:
:
WAVE{SP}{m}{
;
|NL}m
:
0
~
4
Description
:
:
:
:
This command is to set the current C.F. at CC MODE (Peak Value Factor).
This command works only at CC MODE. When BANK varies, these 5 sets of C.F. will
at the same time define different C.F. as shown in Table 4-4. For details, please refer
to 3260A Operation Manual.
For Example
:
:
:
:
a.WAVE 1 ; To set 2nd set C.F.
b.WAVE 4 ; To set 5th set C.F.