Malfunction
Remedy
The display remains empty when the instrument is
1. Check whether the batteries are installed
switched on although the batteries are in place.
with the correct polarity and correct if necessary.
2. If the display is unusual, remove the batteries
and then exchange them for new.
The pressure does not rise although the pump
• Check the connection of the cuff tube and
is running.
connect properly if necessary.
The device frequently fails to measure the
1. Fit the cuff correctly on the arm.
blood pressure values, or the values measured
2. Before starting the measurement make sure that
are too low (too high).
the cuff is not fitted too tightly and that there is
no rolled-up sleeve exerting pressure on the arm
above the measuring position. Take off articles
of clothing if necessary.
3. Measure the blood pressure again in complete
peace and quiet.
Every measurement results in different values
• Please read the following information and
although the device functions normally and
the points listed under «Common sources
the values displayed.
of error». Repeat the measurement.
The blood pressure values measured differ from
• Record the daily development of the values
those measured by the doctor.
and consult your doctor about them.
After the instrument has inflated the cuff the
• Missing plastic ring (distance piece) at
pressure falls very slowly or not at all (no
tube connection. This may only be removed for
reasonable measurement possible).
reasonable measurement possible).
Push this ring over the tube connector so that it
slots in properly and repeat the measurement.
☞
Further information
The level of blood-pressure is subject to fluctuations even with healthy people. Important thereby is,
that
comparable measurements always require the same conditions (Quiet conditions)!
If, in spite of observing all these factors, the fluctuations are larger than 15 mmHg, and/or you hear
irregular pulse tones on several occasions, please consult your doctor.
You must consult your specialist dealer or chemist if there are technical problems with the blood-pressure
instrument.
Never attempt to repair the instrument yourself!
Any unauthorised opening of the instrument invalidates all guarantee claims!