9. Reference to standards
Device standard:
Device corresponds to the requirements of the
European standard for non-invasive blood-pressure
monitor
EN1060-1/ EN1060-3/ EN1060-4
ANSI / AAMI SP10, NIBP – requirements
Electromagn. compatibility:
Device fulfills the stipulations of the European
standard EN 60601-1-2
The stipulations of the EU-Guidelines 93/42/EEC for Medical Products Class IIa have been fulfilled.
10. Technical specifications
Weight:
465 g (with batteries)
Size:
131 (W) x 174 (L) x 73 (H) mm
Storage temperature:
–5 to +50°C
Humidity:
15 to 85% relative humidity maximum
Operation temperature:
10 to 40°C
Display:
LCD-Display (Liquid Crystal Display)
Measuring method:
oscillometric
Pressure sensor:
capacitive
Measuring range:
SYS/DIA:
30 to 280 mmHg
Pulse:
40 to 200 per minute
Cuff pressure display range:
0–299 mmHg
Memory:
Storing the last measurment automatically
Measuring resolution:
1 mmHg
Accuracy:
Pressure within ± 3 mmHg
Pulse ± 5 % of the reading
Power source:
a) 4 x 1.5V batteries; LR6, size AA
b) Mains adapter 6V DC, 600mA (optional)
Cuffs:
Type M for arm circumference 22–32 cm or
Type L for arm circumference 32–42 cm (optional)
Technical alterations reserved!
12