background image

 
M-34-01-06 3/31/2016 

 

 

           

DQ15D/T-DQ15D Manual

 

2 Wire RTD Sensor Calibration (Continued)

 

11 

Press  and  hold  SET  for  1  second.  The  screen 
displays HoLd . WAIT for the display to reset. After 
resetting,  the  connected  precision  resistors’  ap-
proximate  temperature  value  should  display  (i.e. 
511° F or 266° C). 

12 

Turn  OFF  power  and  remove  the  precision  resis-
tors.  Reinstall  the  RTD sensor  and the rear  cover 
of the controller. Return the calibrated controller to 
service. 

 

 

 

3 Wire RTD Sensor Calibration

 

Optional  input  boards,  Item  #  5447  or  5416  need-
ed.  This  board  will accept 2  wire RTDs as well. RTD 
devices are  precision resistors  whose  resistance val-
ue varies with temperature. The connection of  a third 
wire  eliminates  the  natural  resistance  of  the  lead 
wires  to  improve  sensor  accuracy.  The  DQ15D  con-
trol  measures the  RTD resistance (and  the  third  wire 
resistance)  and  compares  that  measurement  with  a 
standard set of  values stored in the memory.

 

You can restore, update or verify that the standard set of 
values is correct.

 

Note:  For  a  1000  ohm  sensor,  the  DIP  switch-
es should be OFF, OFF, OFF. For 100 ohm, ensure 
the  DIP  switches  are  ON,  OFF,  OFF.  See  Dip 
Switch Settings in

 Configuration (Setup).

 

 

Equipment needed

 

 

Two  precision  resistors  (tolerance  +/-  0.1%  or  bet-
ter)  with  a  fixed  value  equal  to  the  RTD  nominal 
value (i.e. 1000 or 100 ohms).

 

 

A  suitable jumper  cable  to  facilitate  changing  input 
resistance.

 

 

A short piece of  jumper wire (simulates third wire).

 

 

 

 

 

 

 

 

Calibration  procedures  require  the  removal  of 
the rear cover of the control. It also requires that 
power  is ON, exposing the technician to poten-
tially  lethal  voltages.  Exercise  EXTREME  CARE 
and  wear  teste

d  electrician’s  gloves  whenever 

power is on. 

 

Les  procédures  de  calibration  nécessitent 
l'enlèvement  du  panneau  arrière  du  contrôle. 
Ça  exige  aussi  que  le  courant  électrique  est 
allumé  ce  qui  expose  le  technicien  à  des  ten-
sions  potentiellement  létales.  Utilisez  toujours 
un  soin  extrême  et  portez  des  gants  certifiés 
d'électricien quand le courrant est allumé.

 

 

Calibration Procedure

 

Turn OFF all power. 

Remove rear cover. 

Remove RTD sensor. 

Install the short piece of jumper wire from terminal 
#1 to #3. 

Install  the  precision  resistors  in  place  of  the  RTD 
sensor, as shown in terminals #2 and #3.  

Install the  jumper  cable  between the loose  end  of 
one of the resistors and the fixed end of the other 
resistor  to  establish  an  input  value  of  a  single  re-
sistor (i.e. 1000 ohms or 100 ohms), as shown.  

Carefully  restore  power  to  the  controller.  Do  not 
come in contact with any exposed voltage. 

 

 

3-Wire RTD Calibration 

– Precision Resistors 

and Jumper Wire in Place 

Summary of Contents for DQ15D

Page 1: ...fset U2 16 OutputSignalOffset U3 16 SignalFilterSetting U4 16 SetPoint1DeadBand U5 16 SetPoint2DeadBand U6 17 PowerSaveSetPointDeadBand U7 17 DisplayStabilizer U8 17 SetPointLimit L 17 SP2andU7Disable...

Page 2: ...Outputs SP1 Set Point reverse acting SPDT 20A resistive 240 VAC max 1HP 240 VAC max 1 2 HP 120VAC SP2 Set Point direct acting SPDT 2A resistive 240 VAC max 1 10HP 240 VAC max 1 20 HP 120VAC ON OFF Dif...

Page 3: ...ys de energize and the screen displays a flashing AAA The power save night setback feature permits the use of a second heating Set Point to conserve energy when required Installation Procedure DQ15D i...

Page 4: ...ill interrupt the heater power supply in the event of low solution level Process Technology heaters include a thermal device Protector 1 2 or 3 on the heater to monitor the heat er s surface temperatu...

Page 5: ...gram Set Points or tem perature limits When DQ15D reaches the Set Point it will energize one or both relays SP1 Heat Set Point for normal Heating Mode opera tion Controls SP1 relay SP2 Cool Set Point...

Page 6: ...he numeric P SET POINT value 4 To change the Set Point value press SET while the P SET POINT value displays value will flash 5 Once the value flashes press or to change the value 6 Press SET once to l...

Page 7: ...r and the fixed end of the other resistor to establish an input value of a single resistor i e 1000 ohms as shown Install the jumper cable between the loose end of one resistor and the fixed end of th...

Page 8: ...Two precision resistors tolerance 0 1 or bet ter with a fixed value equal to the RTD nominal value i e 1000 or 100 ohms A suitable jumper cable to facilitate changing input resistance A short piece of...

Page 9: ...reset After it resets the approximate temperature value for the connected precision resistors should display i e 511 F or 266 C 13 Turn OFF power to the controller and remove the precision resistors...

Page 10: ...settings are OFF ON OFF See Dip Switch Settings in Configura tion Setup The voltage signal must be connected across terminals 1 and 2 of the Adder Board Item 5416 or 5447 Terminal 2 is common negativ...

Page 11: ...rs un soin extr me et portez des gants certifi s d lectricien quand le courrant est allum Calibration Procedure 1 Turn OFF all power 2 Remove rear cover 3 Remove input leads 4 Install the 0 20 mA DC c...

Page 12: ...aussi que le courant lectrique est allum ce qui expose le technicien des ten sions potentiellement l tales Utilisez toujours un soin extr me et portez des gants certifi s d lectricien quand le courran...

Page 13: ...lso requires that power is ON exposing the technician to poten tially lethal voltages Exercise EXTREME CARE and wear tested electrician s gloves whenever power is on Les proc dures de calibration n ce...

Page 14: ...n of the letter c will define the exact nature of the problem c A small c in the upper left hand corner indicates the control is relying on default factory set calibra tion values This happens when th...

Page 15: ...set ting into the view on the screen 2 Press SET to change the value of the setting Once SET has been released the display will flash 3 Use and to scroll through the options for the selected setting...

Page 16: ...ing tells the DQ15D control what type of sensor it is using Value Board Sensor Type Sensor Description 1 5414 or 5416 2 wire RTD Platinum RTD TCR 0 00385 ohm ohm C 2 5414 or 5416 2 wire RTD Platinum R...

Page 17: ...sor and maintained in memory These samples are then averaged to provide an active filter of the signal Using a small value for this setting will cause the DE20 control to respond more quickly to sudde...

Page 18: ...Current Sensor Number represents hundredths of milliamps 0 01 mA Resistance Signal Devices Number represents ohms Frequency Signal Devices Setting represents hertz counts second If the Power Save SET...

Page 19: ...us or Fahrenheit is being displayed If the tem perature being measured is greater than 999 degrees the units are not shown because the display is limited to four positions F5 Temperature Units Convers...

Page 20: ...protection This can happen for example if arc or spot welding equipment is close to the control or if they share a common power line It can occur if flame ignition systems or electrostatic precipitato...