SYSTEM INSTALLATION 

 
The following describes the procedures for 
setting up the system for the first time. 
 

a)

 

Remove all covers, sheets and mattress from the bed. 

 

 

 
 
 

b)

On a standard bed, position Mattress Replacement 

on top of bed, printed top cover facing upwards and 
air  hoses towards the base of the bed. Attach to  the 
bed by securing the adjustable straps under each end 
of  your  bed.  Ensure  buckles  are  securely  fastened 
and straps are pulled tight. On a profiling bed, secure 
the  side  straps  around  the  moveable  sections  of  the 
bed base. 

 
 
 
 
 
 

 
c) 

On a profiling bed, secure the side straps around 

the  moveable  sections  of  the  bed  base. 

DO  NOT 

SECURE  TO  THE  SIDE  RAILS  -  STRAPS 
WILL TEAR OFF. 

 

d) 

Confirm  there  are  no  sharp  objects  in  the 

immediate  area  which  may  risk  damage  to  the 
Mattress Replacement. 

 

 

 
 
 
 

 
Important

Check  that  the  attachment  of  the 

Mattress  Replacement  does  not  interfere  with  the 
movement or operation of the bed. 

 

 
 
 
 

Pump Unit Activation 

 

a)

 

Position Control Unit by hanging hooks over foot 

board of the bed or place unit on the floor under the 
bed. 

 
 
 
 
 
 

b)

 

Attach the air hoses through the use of  the quick 

connector  to  the  Control  Unit.  Ensure  air  hoses  do 
not  kink  between  mattress,  bed  frame  and  Pump 
Unit. 

 

 

 

c)

 

Plug the power cord into an electrical outlet with 

grounded AC power. 

 

NOTE:  Before  inserting  the  plug  into  the  outlet, 
make  sure  the  voltage  is  compatible.  Also  make 
sure this product is well grounded. 

 

d)

 

Switch  the  power  button  on.  Allow  up  to  40 

minutes for full inflation. 

 

e)

 

Once ready, the 

indicator of the power key

 will 

remain on - lay the patient on the bed and select the 
appropriate comfort level. 
 
Perform a “bottom out” test to ensure that patient is 
properly  suspended.  Slide  your  hand  under  the  top 
cover  along  a  deflated  cell  in  the  sacral  (bottom) 
area. Secure sheets loosely enough to ensure they do 
not interfere with cell alternation. 

 

 

 

Summary of Contents for Protekt Aire 4000DX

Page 1: ...Operation Manual FOR Protekt Aire 4000DX 5000DX 270 Washington Street Mount Vernon New York 10553 Phone 1 855 BE PROACTIVE 855 237 7622 Fax 845 205 6005 www proactivemedical com...

Page 2: ...GE CONTENTS 4 PRODUCT FUNCTION 5 Pump 5 Mattress 5 SYSTEM INSTALLATION 6 OPERATION 7 9 General 7 CPR function 9 Pressure set up 9 Low pressure warning 9 CLEANING 10 ENVIRONMENT REQUIREMENTS 11 HANDLIN...

Page 3: ...erly if it has been dropped or damaged or dropped into water Return the product to a service center for examination and repair Keep the cord away from heated surfaces Never block the air openings of t...

Page 4: ...e recommended to use Protekt Aire 4000DX 5000DX system The Protekt Aire 4000DX 5000DX system has been tested and certified for the following standards CE UL Intended Use The Protekt Aire 4000DX 5000DX...

Page 5: ...lowing items should be missing or damaged please contact your local dealer or service center for replacement immediately Pump Unit Mattress Cover Sheet Optional The cover sheet makes cleaning an easy...

Page 6: ...weight Alarm Mute 3 The audible alarm turns on when the mattress pressure is low Press the button to mute alarm Static mode 4 Press to set the air mattress in static therapy mode The function will re...

Page 7: ...hment of the Mattress Replacement does not interfere with the movement or operation of the bed Pump Unit Activation a Position Control Unit by hanging hooks over foot board of the bed or place unit on...

Page 8: ...120 mmHg This may result in cell damage DO NOT change any component by yourself If there is need for replacement or repair always contact your local dealer For patients When the pressure has been adj...

Page 9: ...specific requirements NOTE Press the static mode button from the touch panel to provide a firm surface that makes it easier for the patient to transfer or reposition The static mode can support the pa...

Page 10: ...uggestion from a health care professional Low pressure warning When abnormal pressure occurs the Low Pressure indicator will come on For some types of pumps the alarm will be activated to alert of a l...

Page 11: ...d be wet thoroughly with 1 9 Hypochiorite solution 1 part bleach to 9 parts water and allow drying for at least 10 minutes Then blot with a clean damp cloth Unzip the top cover from the mattress Brush...

Page 12: ...in the pump unit and check the airflow from the hose connection port The airflow should alternate between ports every half cycle time Check the air hoses to see if there are any kinks or breaks For re...

Page 13: ...on PU Pressure range 20 mmHg 50 mmHg Air cells Nylon PVC Cycle time 10 min Base Nylon PVC Case Material Flame Retardant ABS Protekt Aire 4000DX 20 air cells Protekt Aire 5000DX 17 air cells Static Mod...

Reviews: