- 8 - 

NOTE:

 Please follow instructions below for detailed operating procedure of each type.

 

 
Step 1        Turn on the power. The indicator of the power key will flash. 

 

Step 2     

 

The pump will enter into static mode when the power is initially turned on. 

First it will inflate the mattress to 15mmHg, and then enter alternate mode and indicator of 

the power key remains on. 

                   

NOTE: 

                  If 15mmHg still cannot be reached within 40 minutes, the alarm will beep and its LED will 

come  on.

 

Press  the  Alarm  Mute  button  to  mute  the  alarm.  The  alarm  LED  will  continue 

flashing. 

                     
Step  3

   

The  alternating  cycle  time  is  fixed  at  10  minutes.  Weight  range  /pressure  level  is  set  at 

80kg/30mmHg  initially.  Press  the  up/down  buttons  on  panel  to  adjust  the  weight/  pressure 
level to the patient’s specific requirements

.

 

 

                 

NOTE: 

                  Press the static mode button from the touch panel to provide a firm surface that makes 

it  easier  for  the  patient  to  transfer  or  reposition.  The  static  mode  can  support  the 
patient  by  keeping  the  mattress  from  bottoming  out  when  the  patient  is  in  a  sitting 
position.

 

The static mode will revert back to alternating mode after 20 mins.

 

Press the 

Static button again to switch back to alternating mode. 

 
Step  4      During  normal  operation,  the  unit  will  monitor  pressure.  If  the  mattress  pressure  is  lower 

 

than the set pressure, the pump will automatically inflate the mattress to the set level and then 
stop.   

 
NOTE: 

         

In  case  there  is  a  bad  connection  of  tubing  or  any  air  leakage  from  the  mattress,  this  will

 

result in a low-pressure condition, and it will activate the alarm within 1 min. 

 

                  Press the Alarm Mute button to mute the alarm. The alarm LED will continue flashing. 
               

 

Press the Alarm Mute button again to re-enable the audible alarm.

 

 
 
                 

 

               

 

 
 
 
 
 
 

 

 

 
 

Summary of Contents for Protekt Aire 4000DX

Page 1: ...Operation Manual FOR Protekt Aire 4000DX 5000DX 270 Washington Street Mount Vernon New York 10553 Phone 1 855 BE PROACTIVE 855 237 7622 Fax 845 205 6005 www proactivemedical com...

Page 2: ...GE CONTENTS 4 PRODUCT FUNCTION 5 Pump 5 Mattress 5 SYSTEM INSTALLATION 6 OPERATION 7 9 General 7 CPR function 9 Pressure set up 9 Low pressure warning 9 CLEANING 10 ENVIRONMENT REQUIREMENTS 11 HANDLIN...

Page 3: ...erly if it has been dropped or damaged or dropped into water Return the product to a service center for examination and repair Keep the cord away from heated surfaces Never block the air openings of t...

Page 4: ...e recommended to use Protekt Aire 4000DX 5000DX system The Protekt Aire 4000DX 5000DX system has been tested and certified for the following standards CE UL Intended Use The Protekt Aire 4000DX 5000DX...

Page 5: ...lowing items should be missing or damaged please contact your local dealer or service center for replacement immediately Pump Unit Mattress Cover Sheet Optional The cover sheet makes cleaning an easy...

Page 6: ...weight Alarm Mute 3 The audible alarm turns on when the mattress pressure is low Press the button to mute alarm Static mode 4 Press to set the air mattress in static therapy mode The function will re...

Page 7: ...hment of the Mattress Replacement does not interfere with the movement or operation of the bed Pump Unit Activation a Position Control Unit by hanging hooks over foot board of the bed or place unit on...

Page 8: ...120 mmHg This may result in cell damage DO NOT change any component by yourself If there is need for replacement or repair always contact your local dealer For patients When the pressure has been adj...

Page 9: ...specific requirements NOTE Press the static mode button from the touch panel to provide a firm surface that makes it easier for the patient to transfer or reposition The static mode can support the pa...

Page 10: ...uggestion from a health care professional Low pressure warning When abnormal pressure occurs the Low Pressure indicator will come on For some types of pumps the alarm will be activated to alert of a l...

Page 11: ...d be wet thoroughly with 1 9 Hypochiorite solution 1 part bleach to 9 parts water and allow drying for at least 10 minutes Then blot with a clean damp cloth Unzip the top cover from the mattress Brush...

Page 12: ...in the pump unit and check the airflow from the hose connection port The airflow should alternate between ports every half cycle time Check the air hoses to see if there are any kinks or breaks For re...

Page 13: ...on PU Pressure range 20 mmHg 50 mmHg Air cells Nylon PVC Cycle time 10 min Base Nylon PVC Case Material Flame Retardant ABS Protekt Aire 4000DX 20 air cells Protekt Aire 5000DX 17 air cells Static Mod...

Reviews: