- 11 -
ENVIRONMENT REQUIREMENTS
Operating Conditions
Ambient Temperature: 41°F~104°F
Relative Humidity: 15%~90%, non-condensing
Atmospheric pressure: 700hPa to 1060hPa
Storage and shipping conditions
Ambient temperature: -13°F~158°F
Relative Humidity: 10%~90%, non-condensing
Atmospheric pressure: 700hPa to 1060hPa
HANDLING AND STORAGE
Lay the mattress out flat and upside down.
Roll from the foot end towards the head end; the foot-end strap can then be stretched around the
rolled mattress to prevent unrolling.
Do not fold, crease or stack the mattress.
MAINTENANCE
General
Check the power cord and plug to see if there are abrasions or excessive wear.
Check the mattress cover for signs of wear or damage. Ensure the mattress cover and tubes are
connected together correctly.
Plug in the pump unit and check the airflow from the hose connection port. The airflow should
alternate between ports every half-cycle time.
Check the air hoses to see if there are any kinks or breaks. For replacement, please contact your
local agent or dealer.
Make sure the mattress tube is well connected.
Check the pump unit and make sure both power indicators are off when the switch is turned off.
Low pressure
Examine if there is any air leakage between the pump and the mattress connections or from the air
mattress tubes:
Check connectors between the air mattress and pump. If there is any disconnection, please
reconnect it.
Check the CPR Valves. Ensure their outlets are sealed.
Check the air-connecting tubes. Ensure each single cell is not broken.
Set the pressure at Max firm. Keep the tubes fully inflated and inspect for air leakage.
Check if there is any air leakage from cells. Ensure no leakage occurs. If any leakage occurs,
please contact your local agent or dealer.