42
Prior Scientific
5.6 Z Axis Commands
Command Arguments Response
(All end
with <cr>)
Description
BLZH
s,b
0
Sets the Z backlash value for host moves to b in microsteps.
s = 1 enables backlash s=0 disables backlash
BLZH
s
0
Enables / Disables the Z backlash s = 1 enables backlash
s=0 disables backlash.
BLZH
None
s,b
Reports back the Z backlash value for Host moves.
BLZJ
s,b
0
Sets the Z backlash value for Joystick/Digipot to b in microsteps
s = 1 enables backlash s=0 disables backlash.
BLZJ
s
0
Enables / Disables the Z backlash s = 1 enables backlash s=0
disables backlash.
BLZJ
None
s,b
Reports back s and b for Z axis (see above)
C
None
w
Reports the current step size for the focus motor.
C
w
0
Sets the current step size for the focus motor w
D
z
R
Moves Down by z steps.
D
None
R
Moves Down one step as defined by the ‘C’ Command.
FOCUS
None
Text string
Prints information about focus unit..
There are 250 microsteps per full step. The information end is
always a line saying END This allows for the addition of extra
fields of information without effecting application software.
Users should always read lines in until the END is seen to keep
in sync.
Example
FOCUS = NORMAL
TYPE = 0
MICRONS/REV = 100
END
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for OptiScan II ES107
Page 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 92: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 93: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...