Operate a Code Blue Timer
Code Blue Timer operation overview
• Code Blue Timer function is triggered by the input of the wired connection to a code blue system and the use of the
three-button Timer Control Switch allows a user to halt a code blue count event.
• Upon receiving a code blue system input, the code blue timer immediately enters code blue mode and starts counting
up from 00:00:00. Upon reaching 99:59:59, the timer will beep for 3 seconds. If the code blue counter reaches
99:59:59 the counter will stop.
• Code blue mode is indicated by the colon, dash and decimal points not being lit (in both viewing and counting
modes).
• Time duration of the last code blue event is stored in memory, and can be viewed until a new code blue event is
triggered.
• A code blue count can be halted by pressing the STOP/START button. Pressing either DISPLAY or RESET is ignored
while in code blue mode.
• Triggering of the code blue input while a current code blue event is counting is ignored.
• In the event of a power outage, its internal memory maintains the correct time for a minimum of one hour without
power; requires a minimum one hour of charging time.
Code Blue Timer: Timer Switch Control operating specifications
The three-button Timer Switch Control allows a user to halt a code blue event.
Pressing the button...
Active Code Blue Event Mode
Halted Mode
DISPLAY (EXIT)
Holding down DISPLAY (EXIT) for a minimum of
four seconds will display the time of the last
code blue event.
Ignored
Exits code blue mode
and sets timer back to
clock mode.
STOP/START (INCREMENT)
Halts (stops) the code blue
event count at the moment the
button is pressed.
Ignored
RESET (SET/CHANGE)
Ignored
Ignored
Page 33