Frequently Asked Questions
Why does the screen display “Connect electrodes to begin
monitoring”?
1. The monitor is not in study mode and has not recognized your heart
rhythm.
2. Verify the leads are properly connected to the electrodes.
3. Verify the electrodes are securely adhered to your skin without
wrinkles or bubbles. Replace electrodes if necessary.
4. If the display does not proceed to “Monitoring”, disconnect and
reconnect the lead wires in the proper order: black, red, white and
lastly, green.
5.
Press and hold each of the snaps, one by one starting with the
green, to determine if one of the electrodes is not making contact.
−− If pressing on the snap allowed the monitor to display
“Monitoring” replace the electrode and verify the placement
location.
−− Ensure the electrode for the green lead is placed as shown in
the diagram, not too far onto your side or back.
6.
If the message persists, contact 888.500.3522 (press 1, 1, 4)