PRESYS
Instruments
PCON-Y18-LP
Page 13
The state of the output isolation valve is user configurable. To change the state of
the valve, press the
SETTINGS
button shown below and change the
Output
Parameters
(OPEN / CLOSED) in the
ADVANCED
tab.
The valve symbol
appears on the calibrator screen indicating that the Isolation
valve is
OPEN
. This valve is used to isolate the unit under test from the pressure control
module.
Fig. 11
- Output Isolation Valve Status
To zero the current pressure measurement, press the
RESET PRESS
button.
Whenever this function is used, the PCON-Y18-LP executes a
VENT
operation and an
AUTO RESET of the pressure read in the Pressure Control Module and returns to the
control mode.
ATTENTION: Whenever you change the inclination of the PCON-Y18-LP, it is
important to perform the RESET operation of the pressure again.
Fig. 12
- Reset of the current pressure measurement
presys