Page 38
Output Calibration
For the output calibration, follow this procedure for retransmitter output 1.
The retransmitter output can be calibrated by using the Indicator input.
The hardware configuration for the input must be the same chosen for the
output (0 to 5V, 0 to 10V or 0 to 20mA) since the Indicator will measure the output
signal. Therefore, check if the configurations of internal jumpers from the optional
output module and from the CPU board correspond to the same input and output
types.
Make sure that the input type was calibrated previously.
Make the connections listed in table 10 according to the output type to be
calibrated.
Output Type
Output 1 and Input
Terminals
current (0 to 20mA)
voltage (0 to 10V)
terminal 13 (+) and 1 (+)
terminal 14 (-) and 3 (-)
voltage (0 to 5V)
terminal 13 (+) and 2 (+)
terminal 14 (-) and 3 (-)
Table 10 - I/O Terminal Connections for Output Calibration
Enter level 6 of Calibration and select the output to be calibrated. Choose the
type of output (0 to 20mA, 0 to 5V or 0 to 10V) and press ENTER.
The display will show the mnemonic related to the first point of calibration.
There are only two points for output calibration.
For current output, the mnemonics are related to 0 and 20mA electric signals.
For voltage output, the mnemonics are related to 0 and 5V or 0 and 10V signals.
Press ENTER after the display shows the mnemonic related to first or second
point of calibration, so that the display starts to show the output value. Then use the
UP and DOWN keys to set the output value to the electric level indicated by the
mnemonic. After the output value is set, press ENTER. During the first point
calibration (0mA, 0V) be careful not to saturate the output signal.
Finally, return to normal mode by moving down through the levels until
reaching level zero.
Return to factory calibration
The Indicator stores the factory calibration parameter values on the non-
volatile memory, which may be recovered at any time.
In case of a bad performance of the instrument due to an incorrect calibration,
use the REC option.
REC - is the option that allows to recover the factory calibration.
Enter level 6 of Calibration, and choose the INPUT or the OUTPUT
option. Select the REC option and press ENTER in order to recover the values from
factory.