Page 35
Figure 26 shows input and output calibration options in level 6 of Calibration.
Fig. 26 - CALIBRATION level options
Calibration of voltage input (0 to 55mV)
In a 0 to 55mV voltage input calibration connect a voltage source to terminals
2(+) and 3(-). It is required 6 voltage references listed in table 6.
Reference
Mnemonic
0.000 mV
C. 0nV
10.000 mV
C.10nV
20.000 mV
C.20nV
30.000 mV
C.30nV
40.000 mV
C.40nV
50.000 mV
C.50nV