Microx TX3
Operation
The temperature for temperature compensated measurement is displayed either “on-line”, if
the temperature sensor is connected, or “off-line”, if the temperature sensor is not connected.
In the upper right corner warning lights are displayed showing possible errors of the raw data
(amplitude, phase, ambient light).
amplitude:
green: The amplitude is correct.
yellow: The amplitude is critically low; replacement of the sensor is
recommended.
red:
The amplitude is too low; the sensor tip may be damaged or
the sensor cable may not be connected.
phase:
green: The phase angle is within the normal range.
red:
The phase angle is out of limits.
ambient light: green: The ratio of sensor signal to ambient light is acceptable.
red:
The level of background light (e. g. direct sunlight, lamps) is too
high. Reduction of ambient light is recommended.
Clicking on
Hide Raw Values
/
Display Raw Values
will hide or show the raw data values for
amplitude
and
phase
.
Fig. 38
Numerical Display
Fig. 39
Control Bar – warning lights
Fig. 40
Control Bar – warning lights, the raw values are displayed