background image

28 

 

 
 
 
 

© 2011 PreSens Precision Sensing GmbH

Microx TX3 

Operation 

Logging Setup 

Measurements can be done with or without data logging.  

Subsequent data storage is not possible, if you do your measurements without data 

logging.  

If you select 

Measure & Log

, click the 

file location 

button to choose a location for the 

measurement file and name it. In the 

File Description 

area you can add a small note to the 

measurement file which is going to be saved with the file header. The measurement file is 
saved as txt-file in ASCII format. 
 
When all pre-settings are done start the measurement by clicking the 

Start

 button. The 

Calibration 

and 

Data Logging 

tab are hidden now and the selected 

Sampling Rate 

and 

Logging Status 

are displayed in the 

Measurement

 tab.  

 

Press the 

Stop

 button to end the measurement. This will also stop the data logging process. 

  
To close the software select 

File

 / 

Exit

.  

 

4.5.1  Control 

Bar 

 

The control bar contains a numerical display for oxygen and temperature as well as warning 
lights. The control buttons for 

Calibration

, and 

Measurement

 are described in chapters 4.4, 

and 4.5. 

Fig. 37   

Software display – Control Bar 

Fig. 36   

Measurement tab showing sampling rate and  

               logging status 

Summary of Contents for Microx TX3

Page 1: ...Microx TX3 Micro fiber optic oxygen transmitter TRANSMITTERS Instruction Manual...

Page 2: ......

Page 3: ...owners The publisher and the author make no claim to these trademarks While every precaution has been taken in the preparation of this document the publisher and the author assume no responsibility fo...

Page 4: ......

Page 5: ...alibration with Temperature Sensor 20 4 4 2 Calibration without Temperature Sensor 22 4 4 3 Manual Calibration 24 4 5 Measurement 26 4 5 1 Control Bar 28 4 5 2 Graphical Display 30 4 6 Subsequent Data...

Page 6: ......

Page 7: ...ptical oxygen sensors also called optodes have several important features They are small Their signal does not depend on the flow rate of the sample They can be physically divided from the measuring s...

Page 8: ...use For operation a PC notebook is required The Microx TX3 is controlled using a comfortable software which also saves and visualizes the measured values The Microx TX3 has a dual 12 bit analog outpu...

Page 9: ...ture sensor PT 1000 Additionally required equipment not supplied Oxygen sensitive chemical optical microsensor You can find microsensors mounted into different types of housings on www presens de prod...

Page 10: ...green device on orange standby OXYGEN SENSOR ST fiber connector Connect the fiber optic microsensor here TEMP Connector for PT 1000 temperature sensor Connect the PT 1000 temperature sensor for tempe...

Page 11: ...USB RS232 interface male Connect the device with a USB or RS232 data cable to your PC notebook Use the provided parts only CH 1 Analog out channel 1 Connect the device with external devices e g a data...

Page 12: ...r cap should be put back on to keep the optical sensor connector clean It is recommended to clean the ST connector with a dust free cleaning wipe or a cleaning implement for ST connectors before the m...

Page 13: ...lease close all other applications as they may interfere with the software 2 Insert the supplied CD ROM into the respective drive 3 If no dialog opens automatically use the explorer to open the file m...

Page 14: ...Select No not this time from the options and click Next Select Install from a list or specific location Advanced then click Next Select Search for the best driver in these locations and go to Search f...

Page 15: ...bH Microx TX3 Installation Then Windows will copy the required driver files Windows should then display a message indicating that the installation was successful Click Finish to complete the installat...

Page 16: ...oftware is working with English and German regional settings To change the regional settings on your PC press Start and go to the Control Panel Choose Regional and Language Options Select the Regional...

Page 17: ...pace Then press Apply and OK Click Customize again and go to the Date tab now In the drop down menu Short date format you have to select dd MM yy and choose the dot in Date separator Again press Apply...

Page 18: ...USB cable press Start and go to the Control Panel Select System Select the Hardware tab in the System Properties window and click Device Manager You can find the USB serial Port under Ports COM LPT in...

Page 19: ...number you open the Device Manager activate Port COM LPT and double click the USB Serial Port A window opens select the Port Settings tab and click on Advanced Change the COM Port Number to a free po...

Page 20: ...RS232 cable to a serial COM port of your PC notebook Tighten the cable with the screws on your PC notebook 2 Connect the power supply 3 Please close all other applications as they may interfere with...

Page 21: ...choose the right COM port With a right mouse click onto Com Port the dialog Select COM Port opens Select the correct COM port in the drop down menu and click the OK button If no device is detected pl...

Page 22: ...TX3 Operation The window shown below is displayed after the connected transmitter got detected Now connect the oxygen microsensor and temperature sensor to the respective connectors on the transmitter...

Page 23: ...brate without temp sensor the current raw phase values of the oxygen microsensor are measured and stored The temperature value during calibration is set manually in the software and stored This option...

Page 24: ...the currently measured amplitude and phase value of the oxygen microsensor and the temperature are displayed Watch the displayed phase value wait for about 3 minutes until the phase angle is constant...

Page 25: ...es until the phase angle is constant the variation of the phase angle should be smaller than 0 1 and the variation of temperature smaller than 0 1 C and click the Store current value button to the rig...

Page 26: ...de and phase value of the oxygen microsensor are displayed Set the current temperature at the oxygen microsensor by using the up and down arrows or typing in the temperature value Watch the displayed...

Page 27: ...re value Watch the displayed phase value wait for about 3 minutes until the phase angle is constant the variation of the phase angle should be smaller than 0 1 and click the Store current value button...

Page 28: ...eviously measured calibration values for the used oxygen sensor are available you can also use these values for the manual calibration 3 Insert the atmospheric pressure at which the calibration data y...

Page 29: ...and the display in the INFO tab will be updated Abort the calibration by clicking the CANCEL button Data Atmospheric pressure 970 hPa Phase signal Valid range Temperature Valid range Amplitude C C r U...

Page 30: ...performed or the software has been closed after calibration the following message showing the date of the last calibration is displayed Click Continue to go on with the measurement assistant or press...

Page 31: ...ing can be selected This defines the number of averaged measured values The higher the running average the longer the time sampling time used for averaging With a higher running average value set maxi...

Page 32: ...d as txt file in ASCII format When all pre settings are done start the measurement by clicking the Start button The Calibration and Data Logging tab are hidden now and the selected Sampling Rate and L...

Page 33: ...ement of the sensor is recommended red The amplitude is too low the sensor tip may be damaged or the sensor cable may not be connected phase green The phase angle is within the normal range red The ph...

Page 34: ...ing is possible by a left mouse click and movement from the upper left corner to the lower right corner of the area of interest Zooming out is done vice versa or by using the submenu Display Zoom Undo...

Page 35: ...ow information about the software version and some important transmitter settings You can also use the submenu Settings Instrument Info to get to this display If you have any questions concerning your...

Page 36: ...pen Excel on your PC Go to File Open and choose Files of type Text files Choose the measurement file you want to process The Text Import Wizard opens choose Delimited and click Next Then choose semico...

Page 37: ...mode The following rows separated by semicolons list the measuring data The first two rows contain the date and time the third the log time in minutes the fourth the oxygen content in the chosen unit...

Page 38: ...INFO OxyView TX3 V6 02 09 2007 by PreSens GmbH 28 06 2011 14 17 59 MEASURE MODE SETTINGS Dynamic Averaging 4 measure mode 1 sec start time 14 17 59 date DD MM YY time hh mm ss logtime min oxygen airsa...

Page 39: ...ne of the following parameters None Oxygen Phase Amplitude Temperature If you have adjusted the desired settings of the analog outputs and want to connect the transmitter to a data logger please close...

Page 40: ...ng can be avoided and its measurement stability prolonged After changing LED intensity the measurement parameters of the transmitter will change and the sensor has to be recalibrated A warning message...

Page 41: ...axis Undo Zoom restores the original display Clear Charts clears the graphs on the display Dimensions opens a dialog to choose the dimensions for the charts Print Charts prints the charts shown on th...

Page 42: ...or PSt1 Optical connector ST compatible Core Center 100 140 Channels 1 LED peak wavelength 505 nm TEMPERATURE SENSOR Potentiometric temperature sensor PT 1000 Range 0 50 C Resolution 0 1 C Temperature...

Page 43: ...s 8 Stoppbits 1 Parity none Handshake none b USB interface Transmitter port RJ11 4 4 socket Serial interface cable to PC RJ11 4 4 to DSub9 USB interface cable to PC RJ11 4 4 to USB type A transmitter...

Page 44: ...40 2011 PreSens Precision Sensing GmbH Microx TX3 Technical Data OPERATION CONTROL LED at the front panel Red Device off Green Device on Orange Standby DIMENSIONS WEIGHT 185 mm x 110 mm x 45 mm 630 g...

Page 45: ...to choose the parameter O2 Temperature Phase Amplitude 0 4095 V Voltage Output Voltage 0 1000 e g 4095 V Oxygen 0 100 409 5 air sat Ampl 0 20000 81900 r U Phase 0 25 102 375 Temperatur 0 100 409 5 C E...

Page 46: ...t voltage range TTL compatible up to 24 V Trigger mode Low High Low Input must be kept Low for at least 50 s Normal state No current Galvanic isolation 500 V rms Timing Specifications Min rise fall ti...

Page 47: ...pe TR15RA180 which is supplied As an alternative power source a battery can be used that meets the DC input voltage given in chapter 5 Technical Data The battery adapter cable is available as an addit...

Page 48: ...ow Voltage 2006 95 EEC For the evaluation of above mentioned Council Directives following standards were consulted DIN EN 61326 1 2006 10 DIN EN 61010 1 2002 08 Verification of FCC Rules This equipmen...

Page 49: ...th an introduction to work with the Microx TX3 fiber optic oxygen transmitter This manual does not claim to be complete We are endeavored to improve and supplement this version We are looking forward...

Page 50: ......

Page 51: ......

Page 52: ...2011 PreSens Precision Sensing GmbH Manufacturer PreSens Precision Sensing GmbH Josef Engert Stra e 11 93053 Regensburg Germany Phone 49 941 94272100 Fax 49 941 94272111 info PreSens de www PreSens de...

Reviews: