background image

Lubricant must be applied before autoclaving.

10.3 Lubricating

P.10

P.11

P.8

P.9

1. Use a high quality spray lubricant intended for use with 

    dental handpieces / air motors (Not included with device).
2. Insert spray nozzle into drive air tube of air motor (shown 

    in Fig 4). Hold the air motor firmly and spray for 

    approximately 2 to 3 seconds.
3. Lubricant should be seen to be expelled from exhaust air 

    tube. (Fig 5) for Midwest 4 holes.

Fig. 5

Fig. 4

Spray Lubricant

Midwest 4 holes

Midwest 4 holes

1. Use correct eye, face and personal protection to avoid possible infection.

2. Disconnect handpiece from air motor.

3. Remove air motor from air supply hose.

10.1 Preparation

Disconnect the air motor from the tube. Remove gross soiling of the instrument with 

cold water (<40°C) immediately after use. Don’t use a fixating detergent or hot water

(>40°C) as this can cause the fixation of residuals which may influence the result of 

the reprocessing process. Store the instruments in a humid surrounding.

Safe storage and transportation to the reprocessing area to avoid any damage and

contamination to the environment.

10.2 Cleaning

10.2.1 Preparation at the Point of Use:

Do a manual pre-cleaning, until the air motor is visually clean. Submerge the air 

motor in a cleaning solution and flush the lumens with a water jet pistol with cold tap

water for at least 10 seconds. Clean the surfaces with a soft bristle brush.

10.2.2 Pre-cleaning

Use a washer-disinfector meeting the requirements of the ISO 15883 series.

Put the instrument into the machine on a tray. Connect the instrument with the WD 

by using suitable adapter and start the program:

• 4 min pre-washing with cold water (<40°C);

• emptying

• 5 min washing with a mild alkaline cleaner at 55°C

• emptying

• 3 min neutralising with warm water (>40°C);

• emptying

• 5 min intermediate rinsing with warm water (>40°C)

• Emptying

The automated cleaning processes have been validated by using 0.5% neodisher 

MediClean forte (Dr. Weigert).

Note: Acc. to EN ISO 17664 no manual reprocessing methods are required for these 

devices. If a manual reprocessing method has to be used, please validate it prior to use.

10.2.3 Automated Cleaning:

Automated Thermal Disinfection in washer/disinfector under consideration of

national requirements in regards to A0 value (see EN 15883).

A disinfection cycle of 5 min disinfection at 93°C has been validated for this device

to achieve an A0 value of 3000.

10.2.4 Disinfection:

Drying of outside of instrument through drying cycle of washer/disinfector. If needed,

additional manual drying can be performed through lint free towel. Insufflate cavities

of instruments by using sterile compressed air.

10.2.5 Drying:

CAUTION: 

Visual inspection for cleanliness of the air motor. If necessary, p

erform reprocessing process again until the air motor is visibly clean. Before 

packaging and autoclaving, make sure that the air motor has been maintained 

acc. to this instruction.

Summary of Contents for L30

Page 1: ...lowed CAUTION Users are required to protect themselves the patient and third parties from danger Users are responsible for correct and intended usage regular maintenance and continual inspection of the air motor Avoid impact or contamination of air motor Do not disassemble or alter the air motor except as recommended in our user instructions Correct eye face and personal protection must be used by...

Page 2: ...art or contact your authorized Precision Plus dealers if problem persists Model L30 M4 4 holes Hose Connection Type ISO 9168 Type 2 Midwest 4 holes Handpiece Connection Type ISO 3964 Air Supply Pressure 3 kg cm 42 p s i Negative Rotating Speed of Motor 3 kg cm Positive Rotating Speed of Motor 3 kg cm Maximum Speed 20 000 r p m 10 20 000 r p m 10 20 000 r p m 10 Air Consumption 3 kg cm 66NL min Rat...

Page 3: ...ds Clean the surfaces with a soft bristle brush 10 2 2 Pre cleaning Use a washer disinfector meeting the requirements of the ISO 15883 series Put the instrument into the machine on a tray Connect the instrument with the WD by using suitable adapter and start the program 4 min pre washing with cold water 40 C emptying 5 min washing with a mild alkaline cleaner at 55 C emptying 3 min neutralising wi...

Page 4: ...ng blood on outer surface may become clotted and cause operational failure Do not use autoclaves exceeding 137 C during sterilization Sterilize according to EN13060 class B Always follow autoclave manufacturer s instructions for use Reprocessed products should be stored protected from dust in a dry dark and cool place with minimum exposure to germs Autoclave sterilization is the only agreed method...

Page 5: ...sion Plus parts installed Replacement parts will be made available for seven years after the date of discontinuation of the model In order to avoid the health risks to operators handling the disposal of medical devices as well as the risks of environment contamination caused thereof a surgeon or a dentist is required to confirm the device is sterilized prior to disposal Ask a specialized firm who ...

Reviews: