7. Connection & Disconnection of Handpiece to Air Motor

1. Connect handpiece to air motor until it

    locks into place. (Fig.1)

2. Pull on handpiece to make sure it is 

    securely attached to air motor.

1. Hold air motor and handpiece attachment separately and pull the 

    handpiece attachment off in an axial direction.(Fig.1)

7.1 Connection

7.2 Disconnection

P.4

P.6

P.5

P.7

6. Air Supply Pressure Setting

Air supply pressure must be tested and set to comply with specified values in

the table 5. Specifications.

WARNING: 

Do not exceed the specified air supply pressure values in table 

                  5. Specifications.

CAUTION: 

Air supply must be free of dust, moisture and oil.

8. Test Prior to Use

Once handpiece is correctly connected to air motor, check for unusual 

vibration, noise or overheating. If there are any abnormalities, see troubleshooting 

chart or contact your authorized Precision Plus dealers. if problem persists.

Model

L30-M4 (4 holes)

Hose Connection Type

ISO 9168 Type 2

(Midwest 4 holes)

Handpiece Connection Type

ISO 3964

Air Supply Pressure

3 kg/cm² (42 p.s.i.)

Negative Rotating Speed of Motor (3 kg/cm²)

Positive Rotating Speed of Motor (3 kg/cm²)

Maximum Speed

20,000 r.p.m. ±10%

20,000 r.p.m. ±10%
20,000 r.p.m. ±10%

Air Consumption (3 kg/cm²)

66NL/min

Rated Torque

0.0005N.m (at 3 kg/cm²)

5. Specifications

Gear Ratio

1:1

10. Post-Use Maintenance

CAUTION:

Do not use strong/highly acidic water, strong acid/alkaline 

chemicals, chlorine containing solutions or solvents such as 

benzine or thinner to wipe, immerse or clean the device.

Do not use ultrasonic bath during the cleaning process.

CAUTION: 

Motor must have completely stopped rotating before 

                 connecting / disconnecting.

9. Forward/Reverse Rotation

Fig. 2

Fig. 3

1. Forward Rotation:Turn FWD/REV 

    regulating sleeve to “F”.(Fig 2)

    F: Forward Rotation.

2. Reverse Rotation:Turn FWD/REV 

    regulating sleeve to “R”. (Fig 3)

    R: Reverse Rotation.

Fig. 1

Summary of Contents for L30

Page 1: ...lowed CAUTION Users are required to protect themselves the patient and third parties from danger Users are responsible for correct and intended usage regular maintenance and continual inspection of the air motor Avoid impact or contamination of air motor Do not disassemble or alter the air motor except as recommended in our user instructions Correct eye face and personal protection must be used by...

Page 2: ...art or contact your authorized Precision Plus dealers if problem persists Model L30 M4 4 holes Hose Connection Type ISO 9168 Type 2 Midwest 4 holes Handpiece Connection Type ISO 3964 Air Supply Pressure 3 kg cm 42 p s i Negative Rotating Speed of Motor 3 kg cm Positive Rotating Speed of Motor 3 kg cm Maximum Speed 20 000 r p m 10 20 000 r p m 10 20 000 r p m 10 Air Consumption 3 kg cm 66NL min Rat...

Page 3: ...ds Clean the surfaces with a soft bristle brush 10 2 2 Pre cleaning Use a washer disinfector meeting the requirements of the ISO 15883 series Put the instrument into the machine on a tray Connect the instrument with the WD by using suitable adapter and start the program 4 min pre washing with cold water 40 C emptying 5 min washing with a mild alkaline cleaner at 55 C emptying 3 min neutralising wi...

Page 4: ...ng blood on outer surface may become clotted and cause operational failure Do not use autoclaves exceeding 137 C during sterilization Sterilize according to EN13060 class B Always follow autoclave manufacturer s instructions for use Reprocessed products should be stored protected from dust in a dry dark and cool place with minimum exposure to germs Autoclave sterilization is the only agreed method...

Page 5: ...sion Plus parts installed Replacement parts will be made available for seven years after the date of discontinuation of the model In order to avoid the health risks to operators handling the disposal of medical devices as well as the risks of environment contamination caused thereof a surgeon or a dentist is required to confirm the device is sterilized prior to disposal Ask a specialized firm who ...