Air motor should be sterilized after each and every treatment.
1. Air motor should be lubricated as per user manual.
2. Insert the air motor into an autoclave pouch and seal pouch correctly. The
autoclave pouch should comply with the requirements of EN ISO 11607.
3. Sterilize at 134 °C for 5 mins. Sterilization of instruments by applying a fractionated
pre-vacuum steam sterilization process (according to EN 285/EN 13060/EN ISO
17665) under consideration of the respective country requirements. Flash
sterilization is not allowed on lumen instruments!
4. Air motor should be stored in a sealed pouch in a dry, clean and dust free place
out of direct sunlight until required for use.
10.5 Sterilization
10.4 Oil Lubrication
1. Apply 1-2 drops of high quality lubricant oil intended for use with dental
handpiece / air motor (not included with device) into drive air tube (Fig. 1)
2. Connect air motor to air supply hose. Operate the motor for approximately 10 seconds.
P.15
CAUTION
● Always follow the instructions provided by lubricant manufacturer. Be aware of all
precautions and warnings.
● Always hold lubricant can in an upright position.
● Hold air motor firmly to avoid it being dislodged due to lubricant spray pressure.
CAUTION
● Air motor must be cleaned and lubricated before sterilization.
● Any remaining blood on outer surface may become clotted and cause operational
failure.
● Do not use autoclaves exceeding 13
7
°C during sterilization.
● Sterilize according to EN13060, class B. Always follow autoclave manufacturer's
instructions for use.
● Reprocessed products should be stored, protected from dust in a dry, dark and
cool place with minimum exposure to germs.
● Autoclave sterilization is the only agreed method to correctly sterilize this air motor.
The validity of other sterilization methods is not confirmed or guaranteed.
● Resistance to reprocessing: 250 cycles (reprocessing cycles include cleaning and
sterilization
The device has not been subjected to a sterilization process
Serial number
Conforms to CE European Directive of “Medical device directive 93/42/EEC”
Authorized representative in the European Community.
Batch code
Manufacturer
P.12
P.13
P.14
0197
NON
STERILE
Medical Device
Consult instructions for use
Sterilizable in a steam sterilizer (autoclave) at the temperature 134°C
(273°F)
Washer-disinfector for thermal disinfection
12. Symbols
Points to check
Rotation
Process
‘O’ rings
Make sure all O’ rings are in correct place and are undamaged.
Connect and start Air Motor. Check for abnormal rotation, noise, overheating
and vibration.
11. Regular Maintenance Checks
Regular maintenance should be performed every 3 months as per the below chart.
If any abnormalities are found, immediately contact your authorized Precision Plus dealer.
* See trouble shooting chart if problems are discovered.