precisionplus L30 Quick Start Manual Download Page 4

Air motor should be sterilized after each and every treatment.

1. Air motor should be lubricated as per user manual.

2. Insert the air motor into an autoclave pouch and seal pouch correctly. The

autoclave pouch should comply with the requirements of EN ISO 11607.

3. Sterilize at 134 °C for 5 mins. Sterilization of instruments by applying a fractionated

pre-vacuum steam sterilization process (according to EN 285/EN 13060/EN ISO

    17665) under consideration of the respective country requirements. Flash

    sterilization is not allowed on lumen instruments!

4. Air motor should be stored in a sealed pouch in a dry, clean and dust free place

out of direct sunlight until required for use.

10.5 Sterilization

10.4 Oil Lubrication

1. Apply 1-2 drops of high quality lubricant oil intended for use with dental

handpiece / air motor (not included with device) into drive air tube (Fig. 1)

2. Connect air motor to air supply hose. Operate the motor for approximately 10 seconds.

P.15

CAUTION

● Always follow the instructions provided by lubricant manufacturer. Be aware of all

precautions and warnings.

● Always hold lubricant can in an upright position.

● Hold air motor firmly to avoid it being dislodged due to lubricant spray pressure.

CAUTION

● Air motor must be cleaned and lubricated before sterilization.

● Any remaining blood on outer surface may become clotted and cause operational

failure.

● Do not use autoclaves exceeding 13

7

°C during sterilization.

● Sterilize according to EN13060, class B. Always follow autoclave manufacturer's

instructions for use.

● Reprocessed products should be stored, protected from dust in a dry, dark and

cool place with minimum exposure to germs.

● Autoclave sterilization is the only agreed method to correctly sterilize this air motor.

The validity of other sterilization methods is not confirmed or guaranteed.

● Resistance to reprocessing: 250 cycles (reprocessing cycles include cleaning and

sterilization

The device has not been subjected to a sterilization process

Serial number

Conforms to CE European Directive of “Medical device directive 93/42/EEC”

Authorized representative in the European Community.

Batch code

Manufacturer

P.12

P.13

P.14

0197

NON

STERILE

Medical Device

Consult instructions for use

Sterilizable in a steam sterilizer (autoclave) at the temperature 134°C

(273°F)

Washer-disinfector for thermal disinfection

12. Symbols

Points to check
Rotation

Process

‘O’ rings

Make sure all O’ rings are in correct place and are undamaged.

Connect and start Air Motor. Check for abnormal rotation, noise, overheating 

and vibration.

11. Regular Maintenance Checks

Regular maintenance should be performed every 3 months as per the below chart.

If any abnormalities are found, immediately contact your authorized Precision Plus dealer.

* See trouble shooting chart if problems are discovered.

Summary of Contents for L30

Page 1: ...lowed CAUTION Users are required to protect themselves the patient and third parties from danger Users are responsible for correct and intended usage regular maintenance and continual inspection of the air motor Avoid impact or contamination of air motor Do not disassemble or alter the air motor except as recommended in our user instructions Correct eye face and personal protection must be used by...

Page 2: ...art or contact your authorized Precision Plus dealers if problem persists Model L30 M4 4 holes Hose Connection Type ISO 9168 Type 2 Midwest 4 holes Handpiece Connection Type ISO 3964 Air Supply Pressure 3 kg cm 42 p s i Negative Rotating Speed of Motor 3 kg cm Positive Rotating Speed of Motor 3 kg cm Maximum Speed 20 000 r p m 10 20 000 r p m 10 20 000 r p m 10 Air Consumption 3 kg cm 66NL min Rat...

Page 3: ...ds Clean the surfaces with a soft bristle brush 10 2 2 Pre cleaning Use a washer disinfector meeting the requirements of the ISO 15883 series Put the instrument into the machine on a tray Connect the instrument with the WD by using suitable adapter and start the program 4 min pre washing with cold water 40 C emptying 5 min washing with a mild alkaline cleaner at 55 C emptying 3 min neutralising wi...

Page 4: ...ng blood on outer surface may become clotted and cause operational failure Do not use autoclaves exceeding 137 C during sterilization Sterilize according to EN13060 class B Always follow autoclave manufacturer s instructions for use Reprocessed products should be stored protected from dust in a dry dark and cool place with minimum exposure to germs Autoclave sterilization is the only agreed method...

Page 5: ...sion Plus parts installed Replacement parts will be made available for seven years after the date of discontinuation of the model In order to avoid the health risks to operators handling the disposal of medical devices as well as the risks of environment contamination caused thereof a surgeon or a dentist is required to confirm the device is sterilized prior to disposal Ask a specialized firm who ...

Reviews: