7
8
Returns
Returned products require a Return Goods Authorization (RGA) number,
contact Precision Medical, Inc. All returns must be packaged in sealed
containers to prevent damage. Precision Medical, Inc. will not be responsible
for goods damaged in transit. Refer to Precision Medical Inc. Return Policy
available on the Internet, www.precisionmedical.com.
Manuals available on our website; www.precisionmedical.com
NOTE: When returning this Easy Air Compressor, disconnect the power
cord from the back of the device.
Disposal Instructions
This device and its packaging contain no hazardous materials. No special
precautions need to be taken when disposing the device and/or its
packaging.
CAUTION
Excessive dust buildup on filter will reduce performance of Compressor. If
this occurs, clean or replace with a new filter.
The air filter does not filter out saline. Treatments using nebulized saline
will draw the saline mist back into the Compressor. Saline causes rapid
deterioration of the compressor seals and will void the warranty.
WARNING
Turn the Easy Air Compressor off and disconnect from any external power
source before cleaning.
DO NOT attempt to repair. There are no serviceable parts inside.
Please Recycle
Cleaning/Maintenance
Maintenance
The EasyAir Compressor should be serviced if the device does not meet flow
specification.
Return the Easy Air Compressor to your Provider for Compressor
Maintenance (including rebuilding pump).
Contact Precision Medical, Inc. for Preventative Maintenance Services.
Compressor Rebuild Kit - Part #502060C
To clean the filter (weekly):
1. Unplug Compressor.
2. Check air inlet filters for dust buildup.
3. If dust buildup occurs, remove air inlet filters and wash in warm, soapy
water and rinse.
4. Absorb excess water from filter with a dry towel.
5. Return the air inlet filters.
Cleaning for Exterior Components:
Use a clean cloth or sponge dampened with soap and water. Wipe dry with a
clean cloth or paper towel.
Replacement Parts
Part #
Description
503237
Large Filter
502391
Small Filter