Grounding Instructions
The Easy Air Compressor must be plugged into an electrical outlet that is
properly grounded. In the event of an electrical short circuit, grounding re-
duces the risk of electrical shock. This product is equipped with a grounding
circuit.
Component Description
WARNING
Ground reliability can only be achieved when the power cord is connected
to an equivalent receptacle marked “Hospital Grade.”
CAUTION
Missing or illegible labels must be replaced, contact Precision Medical, Inc.
DANGER
Improper use of the grounding plug can result in a risk of electrical shock.
CAUTION
Inspect the Easy Air Compressor for visible damage, DO NOT USE if
damaged.
Operating Instructions
1. Become familiar with the Compressor (see Component Identification).
2. Position the Compressor near a properly grounded electrical receptacle.
Allow a minimum of 4 in (10 cm) clearance on all sides to ensure
adequate air flow for cooling of the Compressor.
3. Ensure the On/Off (I/O) Switch is in the “OFF” (O) position.
4. Ensure the Power Cord is properly connected.
5. Attach appropriate device to the Outlet Fitting securely, without
overtightening.
6. Pull Regulator Knob “OUT” to unlock regulator, then turn Knob fully
counterclockwise. (Failure to do so may cause the device not to start.)
7. Press the On/Off (I/O) Switch to the “ON” (I) position.
8. Adjust pressure/flow.
To INCREASE pressure/flow - Turn Knob CLOCKWISE
To DECREASE pressure/flow - Turn Knob COUNTERCLOCKWISE
9. Push Regulator Knob “IN” to lock pressure/flow setting.
10. Compressor is now ready to use.
11. When the Compressor is not in use, press the On/Off (I/O) Switch to the
“OFF” (O) position.
On/Off (I/O)
Switch
Air
Outlet
Carry
Handle
Gauge
Regulator
Knob
Air Inlet
Filters
Fuse Drawer
AC Inlet
Power Cord
Keeper
Power
Cord