Specifications
Dimensions:
Length: 16 in (40.6 cm)
Width: 7.25 in (18.4 cm)
Height: 10.25 in (26.0 cm)
Weight:
Weight: 19.5 lbs (8.8 kg)
Shipping Weight:
23 lbs (10.4 kg)
Electrical:
PM15-F & PM15-P: 120 VAC ±10%, 60 Hz, 5 Amps
PM15E: 230 VAC ±10%, 50 Hz, 3.2 Amps
PM15EFGB: 220/230 VAC, 50 Hz, 2.8 Amps
Outlet:
Grounded 9/16 - 18 Male Threaded Fitting
Operating Range:
Models Equipped with a Pressure Gauge
Maximum Rated Sustained Pressure: 60 psi
PM15 Output: 50 psi.............................Minimum 13 l/min
PM15E/PM15EFGB Output: 50 psi....Minimum 9 l/min
Models Equipped with a Flow Gauge
PM15-F & PM15EF Output:
(± 1 l/min)
l/min - as set and read on the gauge
Operating
Temperature Range:
50°F to 104°F (10°C to 40°C)
Storage
Temperature Range:
-4°F to 104°F (-20°C to 40°C)
Storage/Operating
Humidity:
Maximum 95% Noncondensing
Safety Systems:
High Temperature Motor Shutdown
Thermal Protector:
Class B, Thermal Limit Automatic Reset
Equipment
Classification:
Class I - with respect to protection from electrical
shock
Type B - degree of protection against electrical shock
IPX0 - degree of protection against ingress of liquids
Mode of Operation - continuous
The Compressor meets the EMC requirements of EN 55011: Group 1 Class B.
The Compressor meets the requirements of CGA V-5: section 5.4.
This Compressor is not suitable for use in the presence of a flammable
anesthetic mixture with air, with oxygen or with nitrous oxide.
Specifications are subject to change without prior notice.
WARNING
To reduce risk of burns, electrocution, fire, or injury to persons:
Close supervision is necessary when this Compressor is used by, on, or
near children or invalids.
Use this Easy Air Compressor only for its “Intended Use” as described in
this manual.
DO NOT use attachments not recommended by Precision Medical, Inc.
DO NOT use the Compressor with any equipment that requires “dry” air,
including ventilators, gas blenders, or flow generators, without a device
to properly dehumidify the discharged air of the PM15.
DO NOT use if the Compressor has a damaged cord or plug.
DO NOT use if the Compressor has been dropped or damaged.
DO NOT drop or insert any object into any openings.
DO NOT cover air vents. Restriction of air flow will cause unit to overheat,
causing the thermal reset to shut down the Compressor.
DANGER
Caution must be exercised when using this Compressor where oxygen is
being administered.
DO NOT use near any type of flame or flammable/explosive substances.
DO NOT allow the Compressor to exceed Operating or Storage
Temperature specifications.
TO REDUCE THE RISK OF ELECTROCUTION
Always unplug the Compressor immediately after using.
DO NOT use while bathing.
DO NOT place or store the Compressor where it can fall or be pulled into
a tub or sink.
DO NOT place or drop into water or other liquid.
DO NOT reach for the Compressor that has fallen into water. Unplug
immediately.
DO NOT attempt to repair the Easy Air Compressor. Any attempt could
result in electrical shock.