4
Oxygen
Analyzer
WARNING
• Use only genuine Precision Medical Inc. accessories and
replacement parts. Failure to do so may seriously impair
the analyzer’s performance. Repair of this device must be
performed by a qualified service technician.
• Calibrate the Oxygen Analyzer before use, or weekly when in
operation, or if environmental conditions change significantly.
(i.e., Elevation, Temperature, Pressure, Humidity.
• Use of the Oxygen Analyzer near devices that generate
electrical fields may cause erratic readings.
• Never autoclave, immerse or expose the Oxygen Analyzer
(including sensor) to high temperatures (>70°C). Never expose
the device to pressure, irradiation vacuum, steam, or chemicals.
• This device does not contain automatic barometric pressure
compensation.
• Although the Sensor of this device has been tested with various
anesthesia gases including nitrous oxide, Halothane, Isoflurane,
Enflurane, Sevoflurane and Desflurane and found to have
acceptably low interference, the device in entirety (including
electronics) is not suitable for use in the presence of a
flammable anesthetic mixture with air or with oxygen or nitrous
oxide. Only the threaded Sensor face, Flow Diverter, and
Tee Adapter may be allowed to contact such a gas mixture.
• NOT for use with inhalation agents. Operating the device
in flammable or explosive atmospheres may result in fire or
explosion.
• No modification of this device is allowed.