14
Oxygen
Analyzer
CAUTION
The Oxygen Sensor should not be used in the presence of
flammable anesthetics such as Diethyl Ether or Cyclopropane.
CLEANING
CAUTION
• DO NOT steam autoclave.
• DO NOT immerse the Oxygen Analyzer into any liquid.
• DO NOT use any strong solvent or abrasive cleaners.
• DO NOT allow any liquid to enter the Oxygen Analyzer or
the Oxygen Sensor; this will damage the Oxygen Analyzer or
Oxygen Sensor and will void the Warranty.
1. Disconnect all connections before cleaning.
2. Clean exterior surfaces of the Oxygen Analyzer, Tee Adapter
and Coiled Cable with a cloth dampened with mild detergent and
water.
3. Wipe dry with a clean cloth.
4. Wipe Oxygen Sensor and Diverter using a lint free dry clean cloth.
MAINTENANCE
Sensor Replacement
Reference “SENSOR INSTALLATION
”
CAUTION
Sensor Replacement must be performed by Qualified Personnel.
Battery Replacement
Replace batteries when “Low Battery” icon
is displayed.
1. Remove 4 screws located on the back and remove cover.
2. Remove and replaced batteries with 2 new AA Alkaline batteries.
Verify correct polarity.
3. Secure back cover onto the Oxygen Analyzer with the 4 screws.