Display Functions and Messages
The meter displays various functions and messages during operation and
programming. The following table shows the various displayed functions and
messages with their description.
Display Parameter
Description/Comments
9999F
Out of range
Indicates the input signal exceeds the full-scale
range of the meter.
CALIb
Calibration
Calibrates meter using a calibrated signal source.
CLEAr
Clear
Clears (resets) maximum or minimum temperatures
reached by the process.
DIA9
Diagnostic
Displays parameter settings one at a time for
diagnostic purposes. Setting can not be changed
under this function.
Error
Error
Indicates calibration was not successful.
F or C
°
F or
°
C
Type of scale to be used Fahrenheit or Celsius.
HI t
High temperature
Displays maximum temperature reached by the
process since last cleared (reset).
InPt 1
Input 1
Sets low input calibration (low input signal must be
applied to the meter).
InPt 2
Input 2
Sets high input calibration (high input signal must
be applied to the meter).
LO t
Low temperature
Displays minimum temperature reached by the
process since last cleared (reset).
LatCk
Latch
Sets alarm set points for latching or non-latching
relay operation.
OFFSET
Input offset
Adds or subtracts a programmed offset value to the
input.
Open
Open
Indicates open sensor circuit.
outPut
Output
Sets the optional 4-20 mA output values.
Set 1
Set point 1
Sets operation for Set point 1.
Set 2
Set point 2
Sets operation for Set point 2.
Set 3
Set point 3
Sets operation for Set point 3.
Set 4
Set point 4
Sets operation for Set point 4.
SetPtS
Set points
Sets alarm set and reset point values.
Setup
Setup
Sets operation of set points for latching or non-
latching mode.
Type
Type
Sets type of thermocouple or RTD to be used as
the input. Jumper JP1 must be configured
accordingly.