Model PD650 Large Display Process Meter
Instruction Manual
35
Set the first calibration point
Input 1 (
INPT 1
) flashes indicating that the meter
is ready to be programmed for the input for the
first calibration point. Press
ENTER
.
Inpt 1
Set the input for the first calibration point
The entire display will flash for three seconds. For
instructions see
Five Basic Digit/Display Setting
Instructions
, page 30.
04.000
then
04.001
final input 1
04.003
Set the display for the first calibration point
Display 1 (
Dspy 1
) flashes indicating that the
meter is ready to be programmed for the display
for the first calibration point. Press
ENTER
.
Dspy 1
Program the display for display 1 (
Dspy 1
) as described
above when the input for the first calibration point was
programmed.
00.000
final display 1
02.500
To quit main menu, calibration, or scaling press
ACK
while displaying main menu
item or while display is flashing input n (InPt
n)
or display n (dspy n), where
n
is
the input or display number.
Minimum Input Span (
Error
Message)
The calibration input signals must be within the range of the meter and input 2
must be greater than input 1. However, the display for input 2 does not have to
be greater than the desired display for input 1.
If input 1 and input 2 signals are too close together an
Error
message will
appear and the display will return to the current input to be calibrated. To
terminate calibration process, press ACK button.
Input Range
Minimum Difference Between Input 1 & Input 2
0-5 V
0.16 V
0-10 V
0.32 V
4-20 mA
1.60 mA