Sabre P PD603 Low-Cost Process Meter
Instruction Manual
4
SPECIFICATIONS
Except where noted all specifications apply to
operation at +25°C.
General
DISPLAY
0.56" (14 mm) high, red LED
Four digits (-1999 to 9999), automatic
lead zero blanking.
DISPLAY
UPDATE RATE
5/second
OVERRANGE
Display flashes
9999
UNDERRANGE
Display flashes
-1999
PROGRAMMING
METHODS
Four front panel buttons
RECALIBRATION
All ranges are calibrated at the factory.
Recalibration is recommended at least
every 12 months.
MAX/MIN
DISPLAY
Max/min readings reached by the
process are stored until reset by the
user or until power to the meter is
turned off.
PASSWORD
Programmable password restricts
modification of programmed settings.
NON-VOLATILE
MEMORY
All programmed settings are stored in
non-volatile memory for a minimum of
ten years if power is lost.
POWER
OPTIONS
85-265 VAC, 50/60 Hz; 90-265 VDC,
20 W max
or 12-36 VDC, 12-24 VAC, 6 W max
FUSE
Required fuse: UL Recognized, 5 A
max, slow blow
Up to 6 meters may share one 5 A fuse
ISOLATED
TRANSMITTER
POWER SUPPLY
P+, P- terminals: 24 VDC
±
10%
@ 200 mA max
NORMAL MODE
REJECTION
64 dB at 50/60 Hz
ISOLATION
4 kV input-to-power line
500 V input-to-24 V supply
OVERVOLTAGE
CATEGORY
Installation Overvoltage Category II:
Local level with smaller transient
overvoltages than Installation
Overvoltage Category III.
ENVIRONMENTAL
Operating temperature range:
0 to 65°C (32 to 150°F)
Storage temperature range:
-40 to 85°C (-40 to 185°F)
Relative humidity:
0 to 90% non-condensing
CONNECTIONS
Removable screw terminal blocks
accept 12 to 22 AWG wire, RJ11 for
factory use only.
ENCLOSURE
1/8 DIN, high impact plastic, UL 94V-0,
color: gray
MOUNTING
1/8 DIN panel cutout required. Two
panel mounting bracket assemblies
provided
TIGHTENING
TORQUE
Screw terminal connectors: 5 lb-in
(0.56 Nm)
OVERALL
DIMENSIONS
2.45" x 4.68" x 4.19"
(62 mm x 119 mm x 106 mm)
(H x W x D)
WEIGHT
8.5 oz (241 g)
WARRANTY
1 year parts & labor. See Warranty
Information and Terms & Conditions on
www.predig.com for complete details.
Process Input
INPUTS
Field selectable:
0-20 mA, 4-20 mA
0-5 V, 1-5 V, 0-10 V, or
±
10 V
ACCURACY
±0.05% FS ±1 count
FUNCTION
Linear
TEMPERATURE
DRIFT
50 PPM/
°
C from 0 to 65
°
C ambient
DECIMAL POINT
Up to three decimal places:
d.ddd
,
dd.dd
,
ddd.d
,
or
dddd
Programmed via front panel buttons
CALIBRATION
RANGE
An
Error
message will appear if input 1
and input 2 signals are too close
together.
Input
Range
Minimum Span
Input 1 & Input 2
4-20 mA
0.40 mA
±
10 V
0.20 V
INPUT
IMPEDANCE
Voltage ranges: greater than 1 M
W
Current ranges: 50 - 100
W
(depending on resettable fuse impedance)
INPUT
OVERLOAD
Current input protected by resettable
fuse. Fuse resets automatically after
fault is removed.