ProtEX-MAX PD8-6060 Explosion-Proof Dual-Input Meter Instruction Manual
30
Multi-Point Linearization (
Linear
)
The process inputs may be calibrated or scaled to any display value within the range of the meter. The
meter is set up at the factory for 2-point linear calibration.
Up to 32 linearization points may be selected. See page 51 for details.
MeterView
®
Pro Software
The meter can also be programmed using the PC-based MeterView Pro software available for free
Data logging for one meter at the time is available with MeterView Pro software. More advanced data
acquisition may be accomplished by using any Modbus RTU compliant software.
In order to program the meter using a computer, the meter must be connected using a USB, RS-232, or
RS-485 serial adapter, see
Ordering
on page 5 for details.
Scaling the Meter without a Signal Source
The process inputs (4-20 mA,
10 VDC) can be scaled to display the process variables in engineering
units.
A signal source is not needed to scale the meter; simply program the inputs and corresponding display
values.
Scaling the Meter for Channel A (
SCal-A
)
Scaling the Meter for Channel B (
SCal-B
)
For instructions on how to program
www.
GlobalTestSupply
.com